multiFiltrate

Service Manual

Technical Manual

Edition: 6/03.07

Part no. M28 003 1

Table of Contents

1 Important Information

1.1 Organization of the Technical Manual …………………………………………………………………………. 1-1

1.2 How to Use the Technical Manual……………………………………………………………………………….. 1-2

1.3 Precautions for Working on the System ……………………………………………………………………… 1-3

1.4 Addresses …………………………………………………………………………………………………………………. 1-4

2 Functional Description

2.1 Extracorporeal Circuit………………………………………………………………………………………………… 2-1

2.1.1 Pumps ……………………………………………………………………………………………………………………….. 2-1

2.1.2 Heaters ………………………………………………………………………………………………………………………. 2-2

2.1.3 Pressure Transducer ……………………………………………………………………………………………………. 2-2

2.1.4 Air Detector and Venous Clamp …………………………………………………………………………………….. 2-2

2.1.5 Optical Detector (Non-Opaque/Opaque Fluid Detector)…………………………………………………….. 2-3

2.1.6 Blood Leak Detector …………………………………………………………………………………………………….. 2-3

2.1.7 Heparin Pump……………………………………………………………………………………………………………… 2-3

2.2 Weighing Units ………………………………………………………………………………………………………….. 2-3

2.3 Ci-Ca Module (Option)………………………………………………………………………………………………… 2-4

2.4 Functional Test (T1 Test) and Error Messages…………………………………………………………….. 2-4

2.4.1 Battery Test, Part 1………………………………………………………………………………………………………. 2-5

2.4.2 Scales Test …………………………………………………………………………………………………………………. 2-5

2.4.3 Pump Test ………………………………………………………………………………………………………………….. 2-6

2.4.4 Pressure Transducer ……………………………………………………………………………………………………. 2-8

2.4.5 Optical Detector…………………………………………………………………………………………………………… 2-9

2.4.6 Air Detector……………………………………………………………………………………………………………….. 2-10

2.4.7 Blood Leak Detector …………………………………………………………………………………………………… 2-11

2.4.8 Heater ………………………………………………………………………………………………………………………. 2-12

2.4.9 Battery Test, Part 2…………………………………………………………………………………………………….. 2-15

2.4.10 Audible Alarm ……………………………………………………………………………………………………………. 2-16

2.4.11 Heparin Pump ……………………………………………………………………………………………………………. 2-17

2.4.12 multiDataLink…………………………………………………………………………………………………………….. 2-18

2.4.13 Ci-Ca Module (Option)………………………………………………………………………………………………… 2-18

2.5 Error Messages………………………………………………………………………………………………………… 2-20

2.5.1 Alarm Messages………………………………………………………………………………………………………… 2-20

2.5.2 Warning Messages…………………………………………………………………………………………………….. 2-22

2.5.3 Fatal Errors ……………………………………………………………………………………………………………….. 2-30

2.5.4 Error Codes of Scales Lo-Level Routines ……………………………………………………………………. 2-33

2.5.5 Error Codes of Pressures Lo-Level Routines……………………………………………………………….. 2-35

2.5.6 Error Codes of Heparin Pump Lo-Level Routines…………………………………………………………. 2-36

2.5.7 Error Codes of Pumps Lo-Level Routines……………………………………………………………………. 2-37

2.6 Overview of Display Identification Numbers ……………………………………………………………… 2-37

Fresenius Medical Care multiFiltrate TM 6/03.07 0-1

3 Installation

3.1 Preface………………………………………………………………………………………………………………………. 3-1

3.2 Important Information on Initial Start-Up …………………………………………………………………….. 3-2

3.3 Initial Start-Up Report multiFiltrate……………………………………………………………………………… 3-3

3.4 Explanations on the Initial Start-Up Report …………………………………………………………………. 3-8

3.5 Installing the Ci-Ca Module (Option)………………………………………………………………………….. 3-16

4 TSC / TMC / Maintenance

4.1 Important Information Regarding the Procedure …………………………………………………………. 4-1

4.2 TSC / MA Report multiFiltrate……………………………………………………………………………………… 4-2

4.3 multiFiltrate TSC Report …………………………………………………………………………………………….. 4-8

4.4 Explanations on the TSC / MA Report ……………………………………………………………………….. 4-12

5 Adjustment Instructions and Tests

5.1 Service Tools …………………………………………………………………………………………………………….. 5-1

5.2 Service Program ………………………………………………………………………………………………………… 5-2

5.2.1 Start …………………………………………………………………………………………………………………………… 5-2

5.2.2 Selecting the Language ………………………………………………………………………………………………… 5-3

5.2.3 System Messages ……………………………………………………………………………………………………….. 5-3

5.2.4 Deactivating and Activating the Heparin Pump ………………………………………………………………… 5-3

5.2.5 Filtrate Bag Monitoring Limit ………………………………………………………………………………………….. 5-4

5.2.6 Dialysate Tubing Arrangement ………………………………………………………………………………………. 5-4

5.2.7 Option MultiDataLink (mDL) ………………………………………………………………………………………….. 5-5

5.2.8 Taring and Calibrating the Scales…………………………………………………………………………………… 5-7

5.2.9 Calibrating the Pressures ……………………………………………………………………………………………… 5-8

5.2.10 Calibrating the Blood Leak Detector ……………………………………………………………………………… 5-10

5.2.11 Verifying the System Values………………………………………………………………………………………… 5-11

5.2.12 Events Memory………………………………………………………………………………………………………….. 5-13

5.2.13 Program Treatment Modes………………………………………………………………………………………….. 5-15

5.2.14 Programming the Ci-Ca Data (Option) ………………………………………………………………………….. 5-16

5.3 Extracorporeal Components …………………………………………………………………………………….. 5-17

5.3.1 Tightness of the Venous Occlusion Clamp…………………………………………………………………….. 5-17

5.3.2 Checking the Pressure Transducers …………………………………………………………………………….. 5-17

5.3.3 Venous Pressure Transducer (P.C.B. LP 450-3)…………………………………………………………….. 5-18

5.3.4 Air Detector (P.C.B. LP 450-3) …………………………………………………………………………………….. 5-20

5.3.5 Optical Detector Sensing Opaque / Non-Opaque Fluid (P.C.B. LP 450-3) …………………………. 5-21

5.3.6 Arterial / Filtrate / PHF Pressure Transducers (P.C.B. LP 343-1) ……………………………………… 5-21

5.3.7 Blood Leak Detector …………………………………………………………………………………………………… 5-22

5.3.8 Setting the Heparin Pump……………………………………………………………………………………………. 5-23

5.4 Programming the Processors …………………………………………………………………………………… 5-25

5.4.1 Display Processor (DP) ………………………………………………………………………………………………. 5-26

5.4.2 Operating and Safety Processors (OP and SP) ……………………………………………………………… 5-26

5.4.3 Interchangeability of Safety and Operating Processor …………………………………………………….. 5-30

5.5 Ci-Ca Module (Option) ………………………………………………………………………………………………. 5-31

0-2 Fresenius Medical Care multiFiltrate TM 6/03.07

5.5.1 Preparing the Functional Test ……………………………………………………………………………………… 5-31

5.5.2 Functional Test of the Drip Counter………………………………………………………………………………. 5-32

5.5.3 Functional Test of the Hall Sensor ……………………………………………………………………………….. 5-32

5.5.4 Functional Test of the Insertion Switch………………………………………………………………………….. 5-32

5.5.5 Functional Test of the Line Occlusion …………………………………………………………………………… 5-33

6 PC Service Software MFT

6.1 Organization of the Quick Guide…………………………………………………………………………………. 6-1

6.2 General Information …………………………………………………………………………………………………… 6-1

6.3 Preparation………………………………………………………………………………………………………………… 6-1

6.3.1 System Requirements ………………………………………………………………………………………………….. 6-1

6.3.2 Software Installation …………………………………………………………………………………………………….. 6-2

6.3.3 Hardware Installation……………………………………………………………………………………………………. 6-3

6.3.4 ServiceCard Description ……………………………………………………………………………………………….. 6-3

6.3.5 Starting the Software ……………………………………………………………………………………………………. 6-3

7 Block Diagrams and Component Layouts

7.1 Block Diagram …………………………………………………………………………………………………………… 7-2

7.2 AC Wiring ………………………………………………………………………………………………………………….. 7-3

7.3 P.C.B. LP 122 Heater Control ……………………………………………………………………………………… 7-4

7.3.1 Block Diagram …………………………………………………………………………………………………………….. 7-4

7.3.2 Description………………………………………………………………………………………………………………….. 7-4

7.3.3 Component Layout ………………………………………………………………………………………………………. 7-5

7.4 P.C.B. LP 123 Pump Control……………………………………………………………………………………….. 7-6

7.4.1 Description………………………………………………………………………………………………………………….. 7-6

7.4.2 Component Layout ………………………………………………………………………………………………………. 7-7

7.5 P.C.B. LP 124 Motherboard ………………………………………………………………………………………… 7-8

7.5.1 Description………………………………………………………………………………………………………………….. 7-8

7.5.2 Component Layout ………………………………………………………………………………………………………. 7-9

7.6 P.C.B. LP 125 Blood Leak Detector …………………………………………………………………………… 7-10

7.6.1 Component Layout …………………………………………………………………………………………………….. 7-10

7.7 P.C.B. LP 127 Scales Board………………………………………………………………………………………. 7-11

7.7.1 Description………………………………………………………………………………………………………………… 7-11

7.7.2 Component Layout …………………………………………………………………………………………………….. 7-11

7.8 P.C.B. LP 128, Power Supply Unit……………………………………………………………………………… 7-12

7.8.1 Block Diagram …………………………………………………………………………………………………………… 7-12

7.8.2 Description………………………………………………………………………………………………………………… 7-12

7.8.3 Component Layout …………………………………………………………………………………………………….. 7-13

7.9 P.C.B. LP 129 User Interface …………………………………………………………………………………….. 7-14

7.9.1 Component Layout …………………………………………………………………………………………………….. 7-14

7.10 P.C.B. LP 244 Operating and Safety Processors………………………………………………………… 7-15

7.10.1 Block Diagram …………………………………………………………………………………………………………… 7-15

7.10.2 Jumper Description…………………………………………………………………………………………………….. 7-15

7.10.3 Component Layout …………………………………………………………………………………………………….. 7-17

Fresenius Medical Care multiFiltrate TM 6/03.07 0-3

7.11 P.C.B LP 343-4 Pressure Transducer ………………………………………………………………………… 7-18

7.11.1 Component Layout …………………………………………………………………………………………………….. 7-18

7.12 P.C.B. LP 450-3 multiFiltrate Air Detector Control ……………………………………………………… 7-19

7.12.1 Component Layout …………………………………………………………………………………………………….. 7-19

7.13 P.C.B. LP 950, Control Board (Heparin Pump)……………………………………………………………. 7-20

7.13.1 Component Layout …………………………………………………………………………………………………….. 7-20

7.14 Ci-Ca Module……………………………………………………………………………………………………………. 7-21

7.14.1 Block Diagram …………………………………………………………………………………………………………… 7-21

7.14.2 Component Layout (P.C.B. LP MS 0407) ………………………………………………………………………. 7-22

0-4 Fresenius Medical Care multiFiltrate TM 6/03.07

1 Important Information

1.1 Organization of the Technical Manual

Page identification Page number 1-3 is to be interpreted as: Chapter 1, page 3.

Editorial information The current edition of this Technical Manual is

6/03.07 = 5th edition, September 2006

In case of updates, the chapters concerned will be replaced.

Refer to the table below to verify that the Technical Manual is up-to­date.

Chapter Current version

1 6/03.07

2 6/03.07

3 6/03.07

4 6/03.07

5 6/03.07

6 6/03.07

7 6/03.07

Chapter 1: Important Information

Changes Manual changes will be released as new editions and supplements. In

general — subject to change without notice.

Fresenius Medical Care multiFiltrate TM 6/03.07 1-1

Chapter 1: Important Information

1.2 How to Use the Technical Manual

Purpose This Technical Manual is intended for service technicians and is to be

used for first studies (to acquire a basic knowledge) and for reference
purposes (for TSC, maintenance and repair). The Technical Manual,
however, does not replace the training courses offered by the
manufacturer.

Requirements Knowledge of the current Operating Instructions for the respective

system.
Background experience in mechanics, electrical and medical
engineering.

Specifications For the specifications of the respective system, refer to the current

Operating Instructions.

Circuit diagrams and
component layouts

Explanation of the Note
and Caution symbols used

The identification on the PCB permits the operator/technician to verify if
the circuit diagram/component layout matches the PCB actually
installed in the system.

Note

Informs the operator that in case of a failure to follow the steps as
described, a specific function will be executed incorrectly or will not be
executed at all, or will not produce the desired effect.

Caution

Advises the operator against certain procedures or actions that could
cause damage to the equipment or may have adverse effects on
operators and patients.

1-2 Fresenius Medical Care multiFiltrate TM 6/03.07

1.3 Precautions for Working on the System

Authorized persons Assembly, extensions, adjustments, modifications or repairs may only

be carried out by the manufacturer or persons authorized by him.

Chapter 1: Important Information

Test equipment and
accessories

Precautions When working on the open system, the following precautions must be

ESD precautions When repairing and when replacing spare parts, observe the applicable

The activities described in the Technical Manual require the availability
of the necessary technical test equipment and accessories.

respected:
 Protect the components against ingress of fluids.
 Do not touch live parts.
 All plugs, connections and components may only be disconnected or

connected if de-energized.

ESD precautions (e.g. EN 100 015-1).

Fresenius Medical Care multiFiltrate TM 6/03.07 1-3

Chapter 1: Important Information

1.4 Addresses

Manufacturer Fresenius Medical Care AG & Co. KGaA

Please address any inquiries to:

D-61346 Bad Homburg
Germany
+49 (0)6172/609-0
www.fmc-ag.com

Service
Central Europe

International
service

Local service

Fresenius Medical Care
Deutschland GmbH
Geschäftsbereich Zentraleuropa
Kundendienst / Servicecenter
Steinmühlstrasse 24b
61352 Bad Homburg
Germany
Phone: +49 6172 609-7100
Fax: +49 6172 609-7102
E-mail: ServicecenterD@fmc-ag.com

Fresenius Medical Care
Deutschland GmbH
Service Support International
Hafenstrasse 9
D-97424 Schweinfurt
Germany
Phone: +49 9721 678-333 (hotline)
Fax: +49 9721 678-130

1-4 Fresenius Medical Care multiFiltrate TM 6/03.07

2 Functional Description

2.1 Extracorporeal Circuit

The elements for maintaining and monitoring the extracorporeal circuit
of the multiFiltrate are as follows:

 Pumps

 Heaters

 Pressure transducer

 Air detector

 Venous clamp

 Non-opaque/opaque fluid detector

 Blood leak detector

 Heparin pump

Chapter 2: Functional Description

2.1.1 Pumps

Altogether, the multiFiltrate is provided with four pumps:

Pump Delivery rate

Blood pump 10  500 ml/min

Filtrate pump 10  180 ml/min

Substituate pump 10  170 ml/min

Dialysate pump 10  170 ml/min

The pumps are driven by direct-current geared motors. To control the
speed, these motors are each provided with a clock pulse generator,
which is directly connected to the motor shaft. The pump processors for
controlling the individual pumps (P.C.B. LP 123) are fitted on the motor
housing.

All pumps are supplied with 24 V. The nominal voltage of the blood
pump motor is 20 V. The nominal voltage of the other pumps, however,
is 24 V.

A Hall sensor in the pump housing and a permanent magnet in each
pump door monitor the state of the pump door.

The line inserting position of the pump rotors for inserting and removing
the pump segments is detected by a combination of a reed switch (in
the pump housing) and a permanent magnet (in the rotor).

The functional test covers a check of all pumps.

Fresenius Medical Care multiFiltrate TM 6/03.07 2-1

Chapter 2: Functional Description

2.1.2 Heaters

2.1.3 Pressure Transducer

To allow heating of replacement fluids, two heaters, which are activated
and monitored independently of each other, are installed in the
multiFiltrate.

The heater foil is applied to the outside of the heater rod and supplied
with approx. 28 V. The temperatures are controlled and monitored by
altogether four NTC sensors (two for the operating processor and two
for the safety processor). This is implemented on P.C.B. LP 122.

Both heaters are tested during the functional test.

Four pressures are measured at the multiFiltrate:

 Arterial pressure,

Measuring transducer (P.C.B. LP 343-1) between the patients
arterial access and the blood pump
Measuring range: 280  300 mmHg

 Pre-hemofilter pressure

Measuring transducer (P.C.B. LP 343-1) between the blood pump
and the filter inlet
Measuring range: 0  750 mmHg

 Venous pressure,

Measuring transducer (P.C.B. LP 450-3) between the filter outlet
and the patients venous access,
Measuring range: 80  500 mmHg.

 Filtrate pressure or dialysate pressure

Measuring transducer (P.C.B. LP 343-1) between the filter
connector and the filtrate pump
Measuring range: 300  300mmHg.

All pressure transducers are subjected to the functional tests.

2.1.4 Air Detector and Venous Clamp

The air detector (P.C.B. LP 450-3) serves for the detection of air in the
extracorporeal blood circuit and operates on the ultrasound principle .
Both the transmitter and the receiver are integrated in the drip chamber
holder. Once the level in the venous drip chamber has fallen below a
certain threshold, the venous clamp is closed. This function is executed
independently of the operating or safety processor. The additional
board AD 28 increases the transmitter voltage during Preparation to
ensure that the level of saline solution is reliably detected.

The air detector is subjected to the functional test.

2-2 Fresenius Medical Care multiFiltrate TM 6/03.07

2.1.5 Optical Detector (Non-Opaque/Opaque Fluid Detector)

The optical detector (P.C.B. LP 450-3) detects, according to the infrared
principle, whether saline solution or blood is present in the tubing
system.

The optical detector is subjected to the functional test.

2.1.6 Blood Leak Detector

The blood leak detector (P.C.B. LP 125) is provided for the detection of
a potential blood loss through the membrane. It is operated applying a
two-color measuring section. In the course of this, red and green light is
alternately transmitted to a reference receiver or, through the filtrate
line, to a measuring receiver.

Both the transmitter and the receiver are integrated in the line holder.

The blood leak detector is subjected to the functional test.

2.1.7 Heparin Pump

Chapter 2: Functional Description

2.2 Weighing Units

The heparin pump is used for continuous heparinization of the blood.

A syringe plunger is moved by means of a carriage bar. The carriage
bar is connected to a threaded spindle via a slide. A microprocessor­controlled stepper motor causes the spindle to rotate. Depending on the
activation, the piston will move up or down.

One Hall sensor each signals when the piston has reached its upper
and lower end of travel. The safety system of the pump comprises a
speed monitoring device (slotted disc with optical sensor) and a motor
current monitoring function. The syringe types are set via a coding
switch (HEX switch).

The weighing units are used for managing the fluid balance during
treatment.

The weighing cells are operated according to the strain gauge principle.
Signal conditioning, including analog-to-digital conversion, is achieved
per weighing cell on the P.C.B. LP 127. The actual weights are
produced by the operating processor.

Scale 1 and scale 2 each have a useful load of approx. 12 kg.

The scales are subjected to the functional test. To test the scales, a test
weight (ball) of defined value must be taken off each scale. Proper
functioning of the scales can be concluded from the correct difference
between the weights before and after lifting.

Fresenius Medical Care multiFiltrate TM 6/03.07 2-3

Chapter 2: Functional Description

2.3 Ci-Ca Module (Option)

The Ci-Ca module is intended for regional citrate anticoagulation in the
CVVHD treatment therapy.

Turning power on The Ci-Ca module requires a supply voltage of 24VDC. This voltage is

provided by the multiFiltrate system’s power supply unit via the
connector in the lower IV pole support of the IV pole located on the right
of the system.
After the multiFiltrate system was turned on by pressing the power
switch on its rear, the supply voltage of the module is connected. The
operating processor of the Ci-Ca module switches into the standby
mode.
If the multiFiltrate systems is then turned on via the I/O key on the front
of the system, the operating processor of the Ci-Ca module will perform
an internal processor test. This test is performed simultaneously to the
processor test of the multiFiltrate system.

The Ci-Ca module communicates with the multiFiltrate basic system via
a serial interface.

Processor test If a test is not passed successfully, the module will not establish

communications with the multiFiltrate system. It is not possible to
perform a treatment with citrate anticoagulation. The multiFiltrate
system recognizes this problem and displays a message which
proposes to use an alternative anticoagulation equipment (e.g. heparin
pump) and which has to be confirmed by the operator.

T1 test The Ci-Ca module performs its own T1 test, independent of the T1 test

of the multiFiltrate system. This test will be started automatically and
simultaneously to the multiFiltrate T1 test after the prompt whether the
starting conditions are met was confirmed with [OK].

This test cannot be skipped or deselected.

System errors If it is still possible, system errors in the Ci-Ca module are shown on the

multiFiltrate display with the indication that citrate anticoagulation is not
available.

2.4 Functional Test (T1 Test) and Error Messages

After it has been turned on the rear and the [I/O] key has been pressed,
the system automatically starts the processor test. After completion of
the processor test, the display test will be performed. In this test, the
numerical characters are represented for 2 seconds in all of the three
fonts used. After this test, the functional test (T1 test) is started
automatically. Depending on the configuration in the SETUP (SETUP
automatic), the test is running in the background or (SETUP detailed)
the test steps are represented separately on the display.

2-4 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

2.4.1 Battery Test, Part 1

At the beginning of the T1 test, the system assumes the terminal voltage
and loads the battery with a defined resistance over the entire test.

2.4.2 Scales Test

In the first step, the temperature compensation and calibration factors
are tested. In the second step, the scales are tared. The linearity test
follows in step three. The 4 test weights are lifted, and the weight (44.3
g) is tested for a tolerance of ±0.5 g. The offset drift is tested in step four.
The test weights are lowered again; the system expects a weight of 0.0
g with a tolerance of ±0.5 g. The time out for each step (2 to 4) is 10 sec.
In the event of an error, messages are emitted. Here, the error
messages section only relates to scales I. The messages are identical
for scales II to IV.

Error message Causes Action required

Scale I

Cal. factor missing, error.
Acknowledge with
[START/RESET] key.

Scale I

Temp. coeff. factors missing, error.
Acknowledge with
[START/RESET] key.

Scale I

Movement detected on Scale I.
Acknowledge with
[START/RESET] key.

Lithium battery on P.C.B. LP 244­OP discharged.

Defect of the EPROM on P.C.B.
LP 127 of the scales specified.

The system is not in a stable
position. The scales are subjected
to a draft. Bags with tubing system
are placed on the scales; this is
especially applicable to filtrate
scales.

The +12 V and/or 12 V voltage(s)
is (are) missing.

Scales mechanics not smooth. Check for smoothness and adjust,

Replace the battery and execute
the calibration step in the Service
program.

Replace the P.C.B. and the
weighing bar and execute the
calibration step in the Service
program.

Set up the system in a stable
position. Remove the system from
the draft, or close the window.
Relieve the scales.

Check the fuses on the PSU board
LP 128 and replace the P.C.B.
LP 128, if necessary.

Caution: All pressures must be
checked and calibrated, if
necessary.

if necessary. Weight check in the
Service program.

Connector not fitted or defective
ribbon cable.

P.C.B. LP 244-OP defective. Replace P.C.B. LP 244-OP.

Fresenius Medical Care multiFiltrate TM 6/03.07 2-5

Fit the connector to the
motherboard and/or to the P.C.B.
LP 127. Replace the ribbon cable.

Caution: All scales and pressures
must be calibrated.

Chapter 2: Functional Description

Error message Causes Action required

Scale I

Test weight outside tolerance.
Acknowledge with
[START/RESET] key.

Scale I

Zero offset outside tolerance.
Acknowledge with
[START/RESET] key.

The system is not in a stable
position. The scales are subjected
to a draft. Bags with tubing system
are placed on the scales; this is
especially applicable to filtrate
scales.

Lifting magnet for ball not
connected or defective.

Ball mechanics not smooth. Check for smoothness and adjust,

The system is not in a stable
position. The scales are subjected
to a draft. Bags with tubing system
are placed on the scales; this is
especially applicable to filtrate
scales.

Scales mechanics not smooth. Check for smoothness and adjust,

Test weight fails to drop back. Ball
mechanics not smooth.

Set up the system in a stable
position. Remove the system from
the draft, or close the window.
Relieve the scales.

Fit the connector to the
motherboard. If defective, the lifting
magnet must be replaced.

if necessary. Weight check in the
Service program.

Set up the system in a stable
position. Remove the system from
the draft, or close the window.
Relieve the scales.

if necessary. Weight check in the
Service program.

Check for smoothness and adjust,
if necessary. Weight check in the
Service program (ESC Service).

2.4.3 Pump Test

In the first step, the stop-by-OP function is tested, i.e. the pumps must
stop running. In this step, the safety processor releases the pumps, and
the operating processor does not activate the pumps. In the second
step, all of the four pumps are activated with a rate of 100 ml/min. The
safety processor checks whether the pumps are running at the correct
rate and whether the reed contact (line inserting position) is actuated. In
the third step, the operating processor activates the pumps and the
safety processor disables the pump activation, i.e. the pumps may not
be running. In the event of an error, messages are emitted. Here, the
error messages section only relates to the blood pump. The messages
are identical for the other pumps, i.e. the filtrate, substituate and
dialysate pumps.

2-6 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

Error message Causes Action required

Blood pump

Blood pump door open.
Acknowledge with
[START/RESET] key.

Blood pump

Not stopped by operating
processor.
Acknowledge with
[START/RESET] key.

Blood pump

Wrong speed, rate.
Acknowledge with
[START/RESET] key.

Blood pump door open. Close the blood pump door.

Magnet in blood pump door
missing.

Hall sensor in pump housing not
connected or defective.

Control electronics (P.C.B. LP 123)
of the specified pump defective.

P.C.B. LP 244-SR defective. Replace P.C.B. LP 244-SR.

Control electronics (P.C.B. LP 123)
of the specified pump defective.

P.C.B. LP 244-SR defective. Replace P.C.B. LP 244-SR.

Clock pulse generator of motor on
P.C.B. LP 123 not connected.

Control electronics (P.C.B. LP 123)
of the specified pump defective.

Replace the blood pump door.

Plug the connector onto the
appropriate P.C.B. LP 123 or
replace the pump housing.

Replace P.C.B. LP 123.

Caution: The BLD must be
calibrated.

Replace P.C.B. LP 123.

Caution: The BLD must be
calibrated.

Connect the clock pulse generator
socket.

Replace P.C.B. LP 123.

Blood pump

Reed contact  line threading
position
Acknowledge with
[START/RESET] key.

Defect of the clock pulse generator
on the geared motor.

Magnet in rotor missing, defective. Replace the rotor.

Reed switch in pump housing not
connected or defective.

Control electronics (P.C.B. LP 123)
of the specified pump defective.

P.C.B. LP 244-SR defective. Replace P.C.B. LP 244-SR.

Replace the geared motor.

Caution: Blood pump
20 V/3300 rpm; all other pumps
24 V/3300 rpm.

Plug the connector onto the
appropriate P.C.B. LP 123 or
replace the pump housing.

Replace P.C.B. LP 123.

Caution: The BLD must be
calibrated.

Fresenius Medical Care multiFiltrate TM 6/03.07 2-7

Chapter 2: Functional Description

Error message Causes Action required

Blood pump

Stop.
Acknowledge with
[START/RESET] key.

Blood pump

Not stopped by safety processor.
Acknowledge with
[START/RESET] key.

2.4.4 Pressure Transducer

+24-V supply voltage missing. Check the fuse on the PSU board

LP 128. If necessary, replace
P.C.B. LP 128.

Caution: All pressures must be
checked and calibrated, if
necessary.

Control electronics (P.C.B. LP 123)
of the specified pump defective.

P.C.B. LP 244-OP defective.
(The pumps are activated serially,
i.e. the operating processor is
defective only if all pumps fail to be
activated.)

Control electronics (P.C.B. LP 123)
of the specified pump defective.

P.C.B. LP 244-SR defective. Replace P.C.B. LP 244-SR.

Replace P.C.B. LP 123.

Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

Replace P.C.B. LP 123.

Caution: The BLD must be
calibrated.

In the first step, the zero point, amplification and detuning factors are
tested. In the second step, the zero points of the pressure transducers
are checked with a tolerance of ±20 mmHg. In the third step, Part and
Pven are detuned to 300 mmHg. The values are checked with a
tolerance of ±20 mmHg. In the fourth step, PPHF and PFil are detuned
to 300 mmHg. The values are checked with a tolerance of ±20 mmHg.
In the event of an error, messages are emitted. The following table only
shows the error messages for «arterial pressure». The messages are
identical for the venous, pre-hemofilter and filtrate pressures.

Error message Causes Action required

Arterial pressure

Factors / Offset lost.
Acknowledge with
[START/RESET] key.

Lithium battery on P.C.B. LP 244­OP discharged.

Replace the battery and execute
the calibration step in the Service
program.

2-8 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

Error message Causes Action required

Arterial pressure

Zero outside tolerance.
Acknowledge with
[START/RESET] key.

Arterial pressure

Detuning outside tolerance
Acknowledge with
[START/RESET] key.

Zero point drifted off. Check the setting and calibrate the

pressure in the Service program.

P.C.B. LP 343-1 (for venous
pressure: P.C.B. LP 450-3)
defective.

P.C.B. LP 244-OP defective. Replace P.C.B. LP 244-OP.

Amplification drifted off. Check the setting and calibrate the

P.C.B. LP 343-1 (for venous
pressure: P.C.B. LP 450-3)
defective.

Replace P.C.B. LP 343-1 (LP 450-

3) and calibrate the pressure in the
Service program.

Caution: When replacing the
P.C.B. LP 450-3, check / adjust the
air detector, optical detector and
venous pressure transducer.

Caution: All scales and pressures
must be calibrated.

pressure in the Service program.

Replace P.C.B. LP 343-1 (LP 450-

3) and calibrate the pressure in the
Service program.

Caution: When replacing the
P.C.B. LP 450-3, check / adjust the
air detector, optical detector and
venous pressure transducer.

2.4.5 Optical Detector

P.C.B. LP 244-OP defective. Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

In the first step, the non-opaque state is checked. In the second step,
the optical detector is detuned and the opaque state is checked. In the
event of an error, messages are emitted.

Fresenius Medical Care multiFiltrate TM 6/03.07 2-9

Chapter 2: Functional Description

Error message Causes Action required

Opt. detector

Senses opaque fluid.
Acknowledge with
[START/RESET] key.

Opt. detector

Fails to sense opaque fluid after
attenuation.
Acknowledge with
[START/RESET] key.

Blood present in the system or
objects inserted in the OD.

OD adjusted improperly. Check the setting of the OD.

P.C.B. LP 450-3 defective. Replace the P.C.B. LP 450-3.

Measuring head of OD defective. Replace and adjust the OD

P.C.B. LP 244-SR defective. Replace P.C.B. LP 244-SR.

OD adjusted improperly. Check the setting of the OD.

P.C.B. LP 450-3 defective. Replace the P.C.B. LP 450-3.

P.C.B. LP 244-OP defective. Replace P.C.B. LP 244-OP.

Remove the blood line or the
objects from the OD.

Caution: When replacing the
P.C.B. LP 450-3, check / adjust the
air detector, optical detector and
venous pressure transducer.

measuring head.

Caution: The BLD must be
calibrated.

Caution: When replacing the
P.C.B. LP 450-3, check / adjust the
air detector, optical detector and
venous pressure transducer.

2.4.6 Air Detector

Caution: All scales and pressures

must be calibrated.

AD 28 defective. Replace AD 28-1.

In the first step, it is checked whether fluid is present in the drip
chamber. if YES, the clamp is activated (opened). In the second step,
the air detector is detuned and the alarm state checked. In the event of
an error, messages are emitted.

2-10 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

Error message Causes Action required

Air detector

LDA1  not in alarm mode.
Acknowledge with
[START/RESET] key.

Air detector

Clamp does not close.
Acknowledge with
[START/RESET] key.

LD adjusted improperly. Check the setting of the LD.

P.C.B. LP 450-3 defective. Replace the P.C.B. LP 450-3.

Caution: When replacing the
P.C.B. LP 450-3, check / adjust the
air detector, optical detector and
venous pressure transducer.

Ultrasonic detector defective. Replace and adjust the ultrasonic

detector.

P.C.B. LP 244-SR defective. Replace P.C.B. LP 244-SR.

Caution: The BLD must be
calibrated.

LD adjusted improperly (LDA2). Check the setting of the LD.

P.C.B. LP 450-3 defective. Replace the P.C.B. LP 450-3.

Caution: When replacing the
P.C.B. LP 450-3, check / adjust the
air detector, optical detector and
venous pressure transducer.

P.C.B. LP 244-SR defective. Replace P.C.B. LP 244-SR.

Caution: The BLD must be
calibrated.

2.4.7 Blood Leak Detector

In the first step, the blood leak detector is checked for being in an
acceptable state. In the second step, the blood leak detector is detuned
and the alarm state checked. In the event of an error, messages are
emitted.

Error message Causes Action required

Blood leak detector

Calibration values incorrect
or missing.
Acknowledge with
[START/RESET] key.

Lithium battery on P.C.B. LP 244­SP discharged.

Replace the battery and execute
the calibration step in the Service
program.

Fresenius Medical Care multiFiltrate TM 6/03.07 2-11

Chapter 2: Functional Description

Error message Causes Action required

Blood leak detector

Outside acceptable range.
Remove filtrate line.
Acknowledge with
[START/RESET] key.

Blood leak detector

Alarm-free after signal attenuation.
Acknowledge with
[START/RESET] key.

2.4.8 Heater

Empty filtrate line or objects
inserted in the BLD sensor head.

P.C.B. LP 125 and/or BLD sensor
head defective.

P.C.B. LP 244-SR defective. Replace P.C.B. LP 244-SR.

P.C.B. LP 125 and/or BLD sensor
head defective.

P.C.B. LP 244-OP defective. Replace P.C.B. LP 244-OP.

In the first step, the safety relay is checked for being in the open state.
In addition, the sensors are checked for interruption and short-circuit. In
the second step, the safety relay is activated and checked for being in
the closed state. In the third step, the bag and foil sensors of OP and
SP are checked for being synchronized. Tolerances are ±1 °C. In the
fourth step, the foil sensor is detuned to >120 °C and, thus, the safety
shutoff mechanism (fire protection) is checked. In the fifth step, the bag
and foil sensors are detuned to >41 °C to check whether the heater is
switched off in case of an overtemperature of the solutions. In the sixth
step, it is checked whether the heater can be activated. In the event of
an error, messages are emitted. Here, the error messages section only
relates to the lower heater. The messages are identical for the upper
heater.

Remove the filtrate line or the
objects.

Replace P.C.B. LP 125 and the
BLD sensor head, and calibrate the
BLD.

Caution: The BLD must be
calibrated.

Replace P.C.B. LP 125 and the
BLD sensor head, and calibrate the
BLD.

Caution: All scales and pressures
must be calibrated.

Error message Causes Action required

Lower heater

Switch-off path SP defective
(Sub. relay not open)
Voltage missing.
Fuse defective.
Acknowledge with
[START/RESET] key.

2-12 Fresenius Medical Care multiFiltrate TM 6/03.07

Safety relay on P.C.B. LP 122
defective.

Activation on P.C.B. LP 244-SR
defective.

Supply voltage for heater not
connected.

Fuse on P.C.B. LP 122 defective. Replace the fuse.

Replace P.C.B. LP 122.

Replace P.C.B. LP 244-SR.

Caution: The BLD must be
calibrated.

Fit the socket on P.C.B. LP 122.

Chapter 2: Functional Description

Error message Causes Action required

Lower heater

Control FET short-circuit
Acknowledge with
[START/RESET] key.

Lower heater

Bag sensor broken.
Acknowledge with
[START/RESET] key.

Lower heater

Foil sensor broken.
Acknowledge with
[START/RESET] key.

Lower heater

Control FET defective. Replace P.C.B. LP 122.

Activation on P.C.B. LP 244-OP
defective.

Socket of sensors not fitted. Fit the socket on P.C.B. LP 122.

Sensor interrupted. Replace the heater.

P.C.B. LP 122 defective Replace P.C.B. LP 122.

Converter on P.C.B. LP 244-SR
defective.

Socket of sensors not fitted. Fit the socket on P.C.B. LP 122.

Sensor interrupted. Replace the heater.

P.C.B. LP 122 defective Replace P.C.B. LP 122.

Converter on P.C.B. LP 244-SR
defective.

Bag sensor short-circuit. Replace the heater.

Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

Replace P.C.B. LP 244-SR.

Caution: The BLD must be
calibrated.

Replace P.C.B. LP 244-SR.

Caution: The BLD must be
calibrated.

Bag sensor short-circuit.
Acknowledge with
[START/RESET] key.

Lower heater

Foil sensor short-circuit.
Acknowledge with
[START/RESET] key.

Lower heater

Sub. relay not closed,
foil defective.
Acknowledge with
[START/RESET] key.

P.C.B. LP 122 defective Replace P.C.B. LP 122.

Converter on P.C.B. LP 244-SR
defective.

Bag sensor short-circuit. Replace the heater.

P.C.B. LP 122 defective Replace P.C.B. LP 122.

Converter on P.C.B. LP 244-SR
defective.

Supply voltage for heater not
connected.

Fuse on P.C.B. LP 122 defective. Replace the fuse.

Heater foil not connected. Connect the heater foil to P.C.B.

Heater foil interrupted. Replace the heater.

Safety relay defective. Replace P.C.B. LP 122.

Activation on P.C.B. LP 244-SR
defective.

Replace P.C.B. LP 244-SR.

Caution: The BLD must be
calibrated.

Replace P.C.B. LP 244-SR.

Caution: The BLD must be
calibrated.

Fit the socket on P.C.B. LP 122.

LP 122.

Replace P.C.B. LP 244-SR.

Caution: The BLD must be
calibrated.

Fresenius Medical Care multiFiltrate TM 6/03.07 2-13

Chapter 2: Functional Description

Error message Causes Action required

Lower heater

Bag sensors,
OP-SP not synchronous
Acknowledge with
[START/RESET] key.

Lower heater

Foil sensors,
OP-SP not synchronous
Acknowledge with
[START/RESET] key.

Heater still warm after last
treatment or Service program.

Sensor board (P.C.B. LP 1220)
defective.

P.C.B. LP 122 defective Replace P.C.B. LP 122.

Converter on P.C.B. LP 244-SR
defective.

Converter on P.C.B. LP 244-OP
defective.

Heater still warm after last
treatment or Service program.

Sensor board (P.C.B. LP 1220)
defective.

P.C.B. LP 122 defective Replace P.C.B. LP 122.

Converter on P.C.B. LP 244-SR
defective.

Allow the heater to cool down.

Replace the heater.

Replace P.C.B. LP 244-SR.

Caution: The BLD must be
calibrated.

Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

Allow the heater to cool down.

Replace the heater.

Replace P.C.B. LP 244-SR.

Caution: The BLD must be
calibrated.

Lower heater

Foil sensor detuning,
no overtemperature cutoff.
Acknowledge with
[START/RESET] key.

Lower heater

Bag sensor detuning,
no overtemperature.
Acknowledge with
[START/RESET] key.

Converter on P.C.B. LP 244-OP
defective.

DIL relay on P.C.B. LP 122
defective.

DIL relay on P.C.B. LP 122
defective.

Activation on P.C.B. LP 244-OP
defective.

Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

Replace P.C.B. LP 122.

Replace P.C.B. LP 122.

Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

2-14 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

Error message Causes Action required

Lower heater

Foil sensor detuning,
no overtemperature.
Acknowledge with
[START/RESET] key.

Lower heater

FET control defective
Acknowledge with
[START/RESET] key.

DIL relay on P.C.B. LP 122
defective.

Activation on P.C.B. LP 244-OP
defective.

FET on P.C.B. LP 122 defective. Replace P.C.B. LP 122.

Activation on P.C.B. LP 44-OP
defective.

Replace P.C.B. LP 122.

Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

2.4.9 Battery Test, Part 2

The first step was carried out at the beginning of the functional test, with
acceptance of the starting value and loading of the battery. In the
second step, the voltage is checked after loading and is compared with
the starting value. In the event of an error, messages are emitted.

Error message Causes Action required

Battery test

Battery not connected, defective.
Acknowledge with
[START/RESET] key.

Battery not connected. Connect the rechargeable battery.

Battery terminal voltage without
load <<18 V.

Converter on P.C.B. LP 244-OP
defective.

Replace the battery.

Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

Battery test

Load test failed.
Acknowledge with
[START/RESET] key.

Fresenius Medical Care multiFiltrate TM 6/03.07 2-15

Resistor or load relay on the PSU,
P.C.B. LP 128, defective.

Converter or activation on P.C.B.
LP 244-OP defective.

Replace P.C.B. LP 128.

Caution: All pressures must be
checked and calibrated, if
necessary.

Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

Chapter 2: Functional Description

Error message Causes Action required

Battery test

Insufficient capacity.
Acknowledge with
[START/RESET] key.

System not used for a prolonged
period or frequent power failure
(preparation).

Battery defective.

Charging circuit on the PSU board,
P.C.B. LP 128, defective.

Converter on P.C.B. LP 244-OP
defective.

Load the battery from the mains for
a minimum of 10 hours.

Replace the battery.

Replace P.C.B. LP 128.

Caution: All pressures must be
checked and calibrated, if
necessary.

Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

2.4.10 Audible Alarm

In the first step, it is checked whether the audible alarm is silenced. In
the second step, the audible alarm is generated and checked for proper
functioning by means of a microphone. In the event of an error,
messages are emitted.

Error message Causes Action required

Audible alarm

Not silenced.
Acknowledge with
[START/RESET] key.

Constant ambient noise. Repeat the test after the noise has

stopped.

Loud humming noise on the
loudspeaker.

Check the route of the loudspeaker
cable. Replace the loudspeaker­microphone unit or the
motherboard P.C.B. LP 124.

The microphone line is disturbed. Check the route of the microphone

cable.

Microphone or amplifier defective. Replace the loudspeaker-

microphone unit or the
motherboard P.C.B. LP 124.

Audible alarm

Not active.
Acknowledge with
[START/RESET] key.

2-16 Fresenius Medical Care multiFiltrate TM 6/03.07

Alarm tone not audible. Loudspeaker cable connected to

P.C.B. LP 124?

Replace the loudspeaker­microphone unit or the
motherboard P.C.B. LP 124.

Error message despite audible
alarm.

Microphone cable connected to
P.C.B. LP 124?

Replace the loudspeaker­microphone unit or the
motherboard P.C.B. LP 124.

Chapter 2: Functional Description

2.4.11 Heparin Pump

After turning power on to the multiFiltrate, a self-test of the heparin
pump will be performed. Immediately before the multiFiltrate self-test,
the operating processor will query the heparin pump status. In the event
of an error, messages are emitted. If the heparin pump is not installed,
not connected to the power supply or deactivated via Service option, no
error message will be displayed.

Error message Causes Action required

Heparin pump

Heparin pump not ready.
No communication.

Heparin pump

Heparin pump optionally
deactivated in Service mode

Heparin pump

Heparin pump detects internal error

Heparin pump

Wrong / unauthorized syringe type
detected

Heparin pump

Unknown hardware error

Heparin pump does not respond. Acknowledge with

[START/RESET] key.

Heparin pump deactivated.

Heparin pump manually
deactivated in the Service mode

Internal heparin pump error without
definite cause.

Wrong syringe type set. Acknowledge with

Heparin pump detects an unknown
hardware error.

Acknowledge with
[START/RESET] key.

Heparin pump deactivated.

Acknowledge with
[START/RESET] key.

Heparin pump deactivated.

[START/RESET] key.

Heparin pump deactivated.

Select a valid syringe type.
0 = 50 ml P syringe

1 = 30 ml heparin syringe
2 = 50 ml Injectomat syringe

Acknowledge with
[START/RESET] key.

Heparin pump deactivated.

Heparin pump

Heparin pump hardware error
Gate array error

Heparin pump

Spike on reset line

Heparin pump

Wrong HEX switch position

Heparin pump

Powerdown

Fresenius Medical Care multiFiltrate TM 6/03.07 2-17

Heparin pump detects an error in
the gate array.

A reset occurred during operation. Acknowledge with

Wrong HEX switch position set. Acknowledge with

Powerdown without 24 V-cutoff. Acknowledge with

Acknowledge with
[START/RESET] key.

Heparin pump deactivated.

[START/RESET] key.

Heparin pump deactivated.

[START/RESET] key.

Heparin pump deactivated.

Set a valid HEX switch position.

[START/RESET] key.

Heparin pump deactivated.

Chapter 2: Functional Description

Error message Causes Action required

Heparin pump

Operating processor internal
communication error with heparin
pump

Heparin pump

Watchdog error

Heparin pump

NOVRAM error

Heparin pump

CAMUS transmission error

2.4.12 multiDataLink

Erroneous data transmission.
Wrong or missing characters.
Check sum error.

The watchdog is not able to
interrupt the 24 V control voltage
for the stepper motor.

The CRC protection of the data
saved in the E2PROM is not
correct.

Not used in the multiFiltrate. Not used in the multiFiltrate.

Immediately before the multiFiltrate self-test, the operating processor
will query the multiDataLink version identifications. A response given by
the mDL is an indicator for its operability. In the event of an error,
messages are emitted. If the multiDataLink is not installed, not
connected to the power supply or deactivated via Service option, no
error message will be displayed.

Acknowledge with
[START/RESET] key.

Heparin pump deactivated.

Acknowledge with
[START/RESET] key.

Heparin pump deactivated.

Acknowledge with
[START/RESET] key.

Heparin pump deactivated.

Error message Causes Action required

multiDataLink

multiDataLink not standby
No communication

multiDataLink

multiDataLink optionally
deactivated in Service mode

multiDataLink does not respond. Acknowledge with

[START/RESET] key.

multiDataLink deactivated.

multiDataLink manually
deactivated in the Service program.

Acknowledge with
[START/RESET] key.

multiDataLink deactivated.

Activate multiDataLink in the
Service program.

2.4.13 Ci-Ca Module (Option)

The test for the Ci-Ca module can not be enabled if the processor test
of the module and the multiFiltrate was not passed successfully. If these
processor tests are performed successfully, the T1 test will be started
automatically as soon as the operator has confirmed that the starting
conditions are met.

The following functions of the Ci-Ca module are tested:

2-18 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

 Stop of the citrate and the calcium pump by the module’s operating

processor
 Stop of the citrate and the calcium pump by the system’s watchdog.
 Function of the citrate pump and the insertion switch
 Function of the calcium pump and the insertion switch

If a test could not be passed successfully, a warning indicating an error
number appears on the multiFiltrate display. The test can be repeated
any number of times, however, it cannot be skipped.

The T1 test of the multiFiltrate system is performed simultaneously to
the T1 test of the Ci-Ca module.

Error
number

281 Lacking Ci Hall impulse after test was started

282 Ci Hall impulse too early during pump stop test

283 Ci Hall impulse too late during pump stop test

285 Ci Hall impulse too early during pump stop test

286 Ci Hall impulse too late during pump stop test

288 Ci Hall impulse too early during insertion switch

289 Ci Hall impulse too late during insertion switch

291 Lacking Ca Hall impulse after test was started

292 Ca Hall impulse too early during pump stop test

Description Possible cause

(>5 s).

(<7 s).

(>9 s).

(<7 s).

(>9 s).

test (<7 s).

test (>9 s).

(>5 s).

(<7 s).

The rotor of the citrate pump is loose, jammed or
blocked.

Insertion switch pressed —> citrate line already
inserted.

The rotor of the citrate pump is loose, jammed or
blocked.

Insertion switch pressed —> citrate line already
inserted during the test.

The rotor of the citrate pump is loose, jammed or
blocked.

Insertion switch pressed —> citrate line already
inserted during the test.

The rotor of the citrate pump is loose, jammed or
blocked.

The rotor of the Ca pump is loose, jammed or
blocked.

Insertion switch pressed —> calcium line already
inserted.

293 Ca Hall impulse too late during pump stop test

(>9 s).

295 Ca Hall impulse too early during pump stop test

(<7 s).

296 Ca Hall impulse too late during pump stop test

(>9 s).

298 Ca Hall impulse too early during insertion switch

test (<7 s).

299 Ca Hall impulse too late during insertion switch

test (>9 s).

Fresenius Medical Care multiFiltrate TM 6/03.07 2-19

The rotor of the Ca pump is loose, jammed or
blocked.

Insertion switch pressed —> calcium line already
inserted during the test.

The rotor of the Ca pump is loose, jammed or
blocked.

Insertion switch pressed —> calcium line already
inserted during the test.

The rotor of the Ca pump is loose, jammed or
blocked.

Chapter 2: Functional Description

2.5 Error Messages

2.5.1 Alarm Messages

These messages cause all pumps to stop

Code Description

E10 Blood pump: Blood pump door open

E11 Blood pump: Stop

E12 Blood pump: Wrong speed, rate

E13 Arterial pressure, arterial pressure too low

E14 Arterial pressure, arterial pressure too high

E15 Venous pressure, venous pressure too low

E16 Venous pressure, venous pressure too high

E17 TMP, TMP too low

E18 TMP, TMP too high

E19 System stopped, Stop key has been pressed

E20 Pre-filter pressure/ Pre-filter pressure too low (for

hemoperfusion)

E21 Pre-filter pressure, pre-filter pressure too low

E22 Pre-filter pressure, pre-filter pressure too high

E23 Non-opaque/opaque fluid detector, venous detector

senses non-opaque or opaque fluid

E24 Air detector, air and/or microbubbles

E25 Blood leak detector, blood leak, hemolysis, membrane

rupture

E26 Blood leak detector, 2 min override active

E27 Power failure, emergency operation available for max.

15 min.

Converted by OP from SP code 130.

E28 End of emergency operation, the unit will turn itself off

automatically after 5 minutes.

Converted by OP from SP code 189.

E29 Pre-filter pressure/ Pre-filter pressure too high (for

hemoperfusion)

E30 Blood lines: Blood line jammed / Hydrophobic filter

P_pHF wet

E31 Blood lines: Rotor broken

2-20 Fresenius Medical Care multiFiltrate TM 6/03.07

Code Description

Chapter 2: Functional Description

E32 Blood lines: Hydrophobic filter P

E33 Blood lines: Hydrophobic filter P

art

ven

wet

wet

E34 Air detector: Transmitting voltage on the air detector too

high

E35 Ci-Ca module: Citrate pump turning too slowly (volume

monitoring)

E36 Ci-Ca module: Citrate pump turning too fast (volume

monitoring)

E37 Ci-Ca module: Calcium pump turning too slowly

(volume monitoring)

E38 Ci-Ca module: Calcium pump turning too fast (volume

monitoring)

E39 Ci-Ca module: Dialysis is started although the primed

Ci/Ca lines are not free from air

E40 Ci-Ca module: Drop counter citrate  insufficient

number of drops; check citrate bag, drip chamber and
tubing system!

E41 Ci-Ca module: Number of drops counted by citrate drop

counter too high; check citrate drip chamber and tubing
system!

E42 Ci-Ca module: Drop counter calcium  insufficient

number of drops; check calcium bag, drip chamber and
tubing system!

E43 Ci-Ca module: Number of drops counted by calcium

drop counter too high; check calcium drip chamber and
tubing system!

E44 Ci-Ca module: Citrate pump turning too slowly; (motor

steps)

E45 Ci-Ca module: Citrate pump turning too fast; (motor

steps)

E46 Ci-Ca module: Calcium pump turning too slowly (motor

steps)

E47 Ci-Ca module: Calcium pump turning too fast (motor

steps)

E48 Ci-Ca module: Citrate insertion switch without

contact,Ci pump segment inserted properly?

E49 Ci-Ca module: Calcium insertion switch without

contact,Ca pump segment inserted properly?

E50 Ci-Ca module: The citrate bag change takes more than

2 min; finish bag change!

Fresenius Medical Care multiFiltrate TM 6/03.07 2-21

Chapter 2: Functional Description

Code Description

E51 Ci-Ca module: The Ca bag change takes more than

2 min; finish bag change!

E52-54 Not used

E55 Ci-Ca module: MPS treatment is permitted for a

maximum of 4 hours!; initiate reinfusion!

E56 Ci-Ca module: HP treatment is permitted for a

maximum of 4 hours!; initiate reinfusion!

E57 Ci-Ca module: Balance switched off; Ca supply

interrupted!

E58 Ci-Ca module: System error of Ci-Ca module:

Anitcoagulation with citrate completed!

E59 Ci-Ca module: Drop counter citrate  insufficient

number of drops; Ci bag change required?

E60 Ci-Ca module: Drop counter calcium  insufficient

number of drops; Ca bag change required?

2.5.2 Warning Messages

These messages cause the balancing to stop. The blood circuit will be
maintained.

For error messages no. 132 through 144 the scales must be
recalibrated.

Code Description

W62 Blood pump: Blood pump door open

W63 Filtrate pump: Cover open

W64 Sub pump: Cover open

W65 Dialysate pump: Cover open

W66 Prime mode: No weight loss on sub-scale-post

(HVCVVH)

W67 Prime mode: No weight loss on sub-scale-pre

(HVCVVH)

W68 Bag change: Overload on sub-scale-post (HVCVVH)

W69 Bag change: Sub-scale post underload (HVCVVH)

W70 Opt. detector: is opaque

W71 Opt. detector: Fails to sense opaque fluid after

attenuation

W72 Scale 1: Temp. coeff. factors missing, incorrect

2-22 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

Code Description

W73 Scale 2: Temp. coeff. factors missing, incorrect

W74 Scale 3: Temp. coeff. factors missing, incorrect

W75 Scale 4: Temp. coeff. factors missing, incorrect

W76 Scale 1: Calibration factor missing, incorrect

W77 Scale 2: Calibration factor missing, incorrect

W78 Scale 3: Calibration factor missing, incorrect

W79 Scale 4: Calibration factor missing, incorrect

W80 Scale 1: Movement detected on Scale 1

W81 Scale 2: Movement detected on Scale 2

W82 Scale 3: Movement detected on Scale 3

W83 Scale 4: Movement detected on Scale 4

W84 Scale 1: Test weight outside tolerance

W85 Scale 2: Test weight outside tolerance

W86 Scale 3: Test weight outside tolerance

W87 Scale 4: Test weight outside tolerance

W88 Scale 1: Test weight fails to drop back

W89 Scale 2: Test weight fails to drop back

W90 Scale 3: Test weight fails to drop back

W91 Scale 4: Test weight fails to drop back

W92 Air detector: LDA1  not in alarm mode

W93 Air detector: Clamp fails to close

W94 Lower heater: Switch-off path SP defective (Sub-relay

not open)

W95 Upper heater: Switch-off path SP defective (dial. relay

not open)

W96 Lower heater: Control FET short-circuit

W97 Upper heater: Control FET short-circuit

W98 Lower heater: Bag sensor broken

W99 Lower heater: Foil sensor broken

W100 Upper heater: Bag sensor broken

W101 Upper heater: Foil sensor broken

W102 Lower heater: Bag sensor short-circuit

W103 Lower heater: Foil sensor short-circuit

Fresenius Medical Care multiFiltrate TM 6/03.07 2-23

Chapter 2: Functional Description

Code Description

W104 Upper heater: Bag sensor short-circuit

W105 Upper heater: Foil sensor short-circuit

W106 Lower heater: Sub. relay not closed, foil defective

W107 Upper heater: Dial. relay not closed, foil defective

W108 Lower heater: Foil sensor detuning, no over-

temperature cutoff

W109 Upper heater: Foil sensor detuning, no over-

temperature cutoff

W110 Lower heater: Bag sensors, OP-SP not synchronous

W111 Upper heater: Bag sensors, OP-SP not synchronous

W112 Lower heater: Bag sensor detuning, no

overtemperature

W113 Lower heater: Foil sensor detuning, no overtemperature

W114 Upper heater: Bag sensor detuning, no

overtemperature

W115 Upper heater: Foil sensor detuning, no overtemperature

W116 Lower heater: FET control defective

W117 Upper heater: FET control defective

W118 Battery test: Battery not connected, defective

W119 Battery test: Load test failed

W120 Battery test: Insufficient capacity

W121 Blood leak detector: Outside acceptable range, remove

filtrate line

W122 Blood leak detector: Alarm-free after attenuation

W123 Prime mode: No weight loss on SUB scale

W124 Prime mode: No weight loss on DIA scale

W125 Prime mode: No weight loss on SUB scale (CVVHDF)

W126 Prime mode: No weight loss on DIA scale (CVVHDF)

W127 Blood leak detector: Calibration values missing,

incorrect

W128 Audible alarm: Not silenced

W129 Audible alarm: not active

W130 Power failure: Emergency operation available for max.

15 minutes (OP generates code 27)

2-24 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

Code Description

W131 Incorrect allocation of scales, repeat Preparation (very

unusual)

W132 Load on scale, incorrect load on scale I (initial weight)

W133 Load on scale, incorrect load on scale II (initial weight)

W134 Load on scale, incorrect load on scale III (initial weight)

W135 Load on scale, incorrect load on scale IV (initial weight)

W136 Balancing, Scale I, leakage from tubing system, objects

W137 Balancing, Scale II, leakage from tubing system, objects

W138 Balancing, upper scales, leakage from tubing system,

objects

W139 Balancing, filtrate bag full, replace or empty filtrate bag

W140 Balancing, filtrate scales moved, leakage from tubing

system

W141 Scale 1, overload or underload, check load, scale

jammed?

W142 Scale 2, overload or underload, check load, scale

jammed?

W143 Scale 3, overload or underload, check load, scale

jammed?

W144 Scale 4, overload or underload, check load, scale

jammed?

W145 Bag change: Overload on SUB scale (CVVHDF)

W146 Bag change: Overload on SUB scale 1 (CVVH, HF)

W147 Bag change: Overload on SUB scale 2 (CVVH, HF)

W148 Bag change: SUB on wrong scale (CVVHDF)

W149 Bag change: Insufficient substituate (CVVH, HF)

W150 Bag change: Insufficient substituate (CVVHDF)

W151 Bag change: Insufficient saline solution on scale 1

(MPS)

W152 Bag change: Overload on plasma scale 1 (MPS)

W153 Bag change: Overload on plasma scale 2 (MPS)

W154 Bag change: Overload on dialysate scale (CVVHDF)

W155 Bag change: Overload on dialysate scale 1 (CVVHD)

W156 Bag change: Overload on dialysate scale 2 (CVVHD)

W157 Dialysate on wrong scale (tubing arrangement test after

bag change HDF)

Fresenius Medical Care multiFiltrate TM 6/03.07 2-25

Chapter 2: Functional Description

Code Description

W158 Bag change: Insufficient dialysate (CVVHD)

W159 Bag change: Insufficient dialysate (CVVHDF)

W160 Bag change: Overload on filtrate scale

W161 TMP/MPS TMP >100mmHg

W162 Lower heater: MPS temperature > 37°C

W163 Upper heater: MPS temperature > 37°C

W164 Lower heater: Temperature too high

W165 Upper heater: Temperature too high

W166 Bag change: Overload SUB pre (HVCVVH)

W167 Bag change: Insufficient SUB pre (HVCVVH)

W168 Balancing: Ultrafiltration or substitution rate too high

W169 Lower heater: Foil sensors, OP-SP not synchronous

W170 Upper heater: Foil sensors, OP-SP not synchronous

W171 Balancing: Aborted because maximum allowable

deviation of 500g was exceeded

W172 Lower heater: MPS foil temperature too high, risk of

hemolysis

W173 Upper heater: MPS foil temperature too high, risk of

hemolysis

W174 Balancing: Balancing stopped, balancing was switched

off

W175 Battery capacity: Charging voltage too high, battery

deactivated, no power failure

W176 Filtrate scale: Test weight outside tolerance (T0 test)

W177 Filtrate scale: Test weight fails to drop back (T0 test)

W178 Balancing: UF goal reached, ultrafiltration was set to 0

W179 Balancing: Substituate bag empty, replace substituate

bag

W180 Balancing: Dialysate bag empty, replace dialysate bag

W181 Balancing: Movement detected on filtrate scales,

leakage from tubing system

W182 Filtrate scale: Overload or underload, check load, scale

jammed?

W183 Balancing: Plasma bag empty, open outlet line(s) on

plasma bags

W184 Balancing: Plasma pump delivers from wrong scale (I)

2-26 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

Code Description

W185 Balancing: Plasma pump delivers from wrong scale (II)

W186 Balancing: Sub-postdilution bag empty, replace

substituate bag

W187 Balancing: Sub-predilution bag empty, replace

substituate bag

W188 End of treatment: Plasma infused (MPS only)

W189 End of emergency operation: The unit will turn itself off

automatically after 5 minutes

W190 Power failure end

W191 Bag change: Insufficient plasma on scale 2

W192 Prime mode: Overload on dialysate scale (CVVHDF)

W193 Prime mode: Overload on SUB scale (CVVHDF)

W194 Prime mode: Insufficient substituate (CVVHDF)

W195 Prime mode: Insufficient dialysate (CVVHDF)

W196 Prime mode: Overload on sub-scale-pre (HVCVVH)

W197 Prime mode: Overload on sub-scale-post (HVCVVH)

W198 Prime mode: Insufficient SUB pre (HVCVVH)

W199 Prime mode: Insufficient SUB post (HVCVVH)

W200 Blood pump: Not stopped by operating processor

W201 Filtrate pump: Not stopped by operating processor

W202 Sub pump: Not stopped by operating processor

W203 Dialysate pump: Not stopped by operating processor

W204 Blood pump: Wrong speed, rate

W205 Filtrate pump: Wrong speed, rate

W206 Sub pump: Wrong speed, rate

W207 Dialysate pump: Wrong speed, rate

W208 Blood pump: Reed contact  line threading position

W209 Filtrate pump: Reed contact  line threading position

W210 Sub pump: Reed contact  line threading position

W211 Dialysate pump: Reed contact  line threading position

W212 Blood pump: Stop

W213 Filtrate pump: Stop

W214 Sub pump: Stop

Fresenius Medical Care multiFiltrate TM 6/03.07 2-27

Chapter 2: Functional Description

Code Description

W215 Dialysate pump: Stop

W216 Blood pump: Not stopped by safety processor

W217 Filtrate pump: Not stopped by safety processor

W218 Sub pump: Not stopped by safety processor

W219 Dialysate pump: Not stopped by safety processor

W220 Arterial pressure: Factors / offset missing, incorrect

W221 Venous pressure: Factors / offset missing, incorrect

W222 PHF pressure: Factors / offset missing, incorrect

W223 Filtrate pressure: Factors / offset missing, incorrect

W224 Arterial pressure: Zero outside tolerance

W225 Venous pressure: Zero outside tolerance

W226 PHF pressure: Zero outside tolerance

W227 Filtrate pressure: Zero outside tolerance

W228 Arterial pressure: Detuning outside tolerance

W229 Venous pressure: Detuning outside tolerance

W230 PHF pressure: Detuning outside tolerance

W231 Filtrate pressure: Detuning outside tolerance

W232 Heparin pump: does not respond

W233 Heparin pump: manually deactivated in Service

program

W234 Heparin pump: detects an internal problem

W235 Heparin pump: Wrong/invalid syringe type

W236 Heparin pump: No anticoagulation

W237 Heparin pump: Anticoagulation stopped

W238 Heparin pump: Anticoagulant Used Up

W239 Heparin pump: Syringe not changed

W240 Heparin pump: Non-registered error code

W241 Heparin pump: Gate-array error

W242 Heparin pump: Restart after spike reset

W243 Heparin pump: Hex switch position error

W244 Heparin pump: Powerdown without 24V cutoff

W245 Heparin pump: Slotted disc error

2-28 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

Code Description

W246 Heparin pump: Slotted disc not found

W247 Heparin pump: Communication sumcheck error

W248 Heparin pump: No heparin pump STOP

W249 Heparin pump: Communication error with OP

W250 Heparin pump: Watchdog error

W251 Heparin pump: NOVRAM error

W252 CAMUS transmission error

W253 Not used

W254 Not used

W255 mDL not ready/no communication

W256 mDL deactivated via Service option

W257 The BP rate is < 30 ml/min; no rotor breakage

monitoring; closure of pressure transducer is not
detected.

W259 Lower heater: Switch-off path SP defective (Sub-relay

not open), voltage not present, SI defective (for P.C.B.
LP122-B)

W260 Upper heater: Switch-off path SP defective (dial. relay

not open), voltage not present, SI defective (for P.C.B.
LP122-B)

W261 Lower heater: Control FET short circuit (for P.C.B.

LP122-B)

W262 Upper heater: Control FET short circuit (for P.C.B.

LP122-B)

W263 Air detector: Transmitting voltage for air detector not

present / nonconforming (for AD28)

W264 Ci-Ca module: Ci-Ca module does not respond, select

alternate anticoagulation if necessary!

W265 Ci-Ca module: Switch-off path «sleep» does not react,

select alternate anticoagulation if necessary!

W266 Balancing: HF bag change aborted

W281 Ci-Ca module: Citrate pump — test F1, Rotor slack or

stuck?

W282 Ci-Ca module: Citrate pump — OP test F2, rotor slack?

W283 Ci-Ca module: Citrate pump — OP test F3, rotor slack or

stuck?

W285 Ci-Ca module: Citrate pump — WD test F5, rotor slack?

Fresenius Medical Care multiFiltrate TM 6/03.07 2-29

Chapter 2: Functional Description

Code Description

W286 Ci-Ca module: Citrate pump — WD test F6, rotor slack or

stuck?

W288 Ci-Ca module: Citrate pump — insertion switch test F8,

line pump segment inserted too early or rotor slack?

W289 Ci-Ca module: Citrate pump — insertion switch test F9,

rotor slack or stuck?

W291 Ci-Ca module: Calcium pump test F11, rotor slack or

stuck?

W292 Ci-Ca module: Calcium pump — OP test F12, rotor

slack?

W293 Ci-Ca module: Calcium pump — OP test F13, rotor slack

or stuck?

W295 Ci-Ca module: Calcium pump — WD test F15, rotor

slack?

W296 Ci-Ca module: Calcium pump — WD test F16, rotor slack

or stuck?

2.5.3 Fatal Errors

W298 Ci-Ca module: Calcium pump — insertion switch test

F18, line pump segment inserted too early or rotor
slack?

W299 Ci-Ca module: Calcium pump — insertion switch test

F19, rotor slack or stuck?

W330 Ci-Ca module: Balance switched off!, Ca supply

interrupted!

W331 Ci-Ca module: Filtrate rate too low for set Ca dose!,

Increase dialysate or substituate or Ca dose if
necessary!

W332 Ci-Ca module: Filtrate rate too high for set Ca dose!,

Reduce dialysate or substituate or Ca dose if
necessary!

W333 Ci-Ca module: Blood flow too low for set citrate dose!,

Increase blood flow or citrate dose if necessary!

W334 Ci-Ca module: Blood flow too high for set citrate dose!,

Reduce blood flow or citrate dose if necessary!

These messages cause the treatment to stop.

2-30 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

Code Description

E500 Scale 1: (factors missing during treatment)

E501 Scale 2: (factors missing during treatment)

E502 Scale 3: (factors missing during treatment)

E503 Scale 4: (factors missing during treatment)

E504 Memory error: (double values incorrect)

E505 Blood pump: (does not respond)

E506 Substituate pump: (does not respond)

E507 Dialysate pump: (does not respond)

E508 Filtrate pump: (does not respond)

E509 Overvoltage +5V

E510 Overvoltage +12V

E511 Processor test OP: CPU defective,

E512 Processor test OP: Incorrect CRC interrupt vectors

E513 Processor test OP: CRC CRC area defective

E514 Processor test OP: CRC Code area defective

E515 Processor test OP: RAM, bank B defective

E516 Processor test OP: RAM, bank 8 defective

E517 State transition

E518 Program sequence

E519 Self-Test

E520 Balancing

E521 SP does not respond

E522 DP does not respond

E523 Program flow OP: Watchdog

E524 Processor test OP: Watchdog does not respond

E525 Processor test SP: CPU

E526 Processor test SP: Program memory  vectors

E527 Processor test SP: Program memory — CRC area

E528 Processor test SP: Program memory — code area

E529 Processor test SP: RAM — bank B

E530 Processor test SP: RAM — bank B

E531 Processor test SP: Watchdog does not respond

Fresenius Medical Care multiFiltrate TM 6/03.07 2-31

Chapter 2: Functional Description

Code Description

E532 Invalid software combination: OP, DP, SP

E533 Ci-Ca module: Logoff of the option by the MFT

E534 Ci-Ca module: Alarm of the option not followed by MFT

alarm

E535 Ci-Ca module: Timeout of the status message of the

MFT (>2.5 s)

E536 Ci-Ca module: Too many status messages of the MFT

(2x<0.5 s)

E537 Ci-Ca module: Receipt of message not confirmed

E538 Ci-Ca module: unknown identifier

E539 Ci-Ca module: Write buffer full

E540 Ci-Ca module: Read buffer full

E541 Ci-Ca module: Unable to write on TXe

E542 Ci-Ca module: Missing setup values from the MFT (at

end of line installation)

E550 Ci-Ca module: Too many extra Hall pulses from the Ca

pump

E551 Ci-Ca module: Ca pump could not be stopped

E560 Ci-Ca module: Too many extra Hall pulses from the

citrate pump

E561 Ci-Ca module: Citrate pump could not be stopped

E570 Ci-Ca module: Error on checking variable replication

(Int)

E571 Ci-Ca module: Error on checking variable replication

(Float)

E572 Ci-Ca module: Error on checking variable replication

(Long)

E573 Ci-Ca module: Error on checking software diversity

2-32 Fresenius Medical Care multiFiltrate TM 6/03.07

2.5.4 Error Codes of Scales Lo-Level Routines

Scale no. 1

Code Description

E30001 Standstill not possible (timeout after 10 sec)

E30002 Weight value out of range (overload or underload)

E30003 Incorrect calibration weight value

E30004 Negative initial weight

E30005 Wrong initial weight

E30006 ACT_WGT 65535

E30007 Alarm limits permanently violated

E30008 INF_WGT invalid

E30050 Calibration factor missing or incorrect

Chapter 2: Functional Description

Scale no. 2

E30051 Temp. coeff. factors missing or incorrect

E30052 Double fuse defective (zero value)

E30053 Temperature sensor missing / defective

Code Description

E30101 Standstill not possible (timeout after 10 sec)

E30102 Weight value out of range (overload or underload)

E30103 Incorrect calibration weight value

E30104 Negative initial weight

E30105 Wrong initial weight

E30106 ACT_WGT 65535

E30107 Alarm limits permanently violated

E30108 INF_WGT invalid

E30150 Calibration factor missing or incorrect

E30151 Temp. coeff. factors missing or incorrect

E30152 Double fuse defective (zero value)

E30153 Temperature sensor missing / defective

Fresenius Medical Care multiFiltrate TM 6/03.07 2-33

Chapter 2: Functional Description

Scale no. 3

Code Description

E30201 Standstill not possible (timeout after 10 sec)

E30202 Weight value out of range (overload or underload)

E30203 Incorrect calibration weight value

E30204 Negative initial weight

E30205 Wrong initial weight

E30206 ACT_WGT 65535

E30207 Alarm limits permanently violated

E30208 INF_WGT invalid

E30250 Calibration factor missing or incorrect

E30251 Temp. coeff. factors missing or incorrect

E30252 Double fuse defective (zero value)

Scale no. 4

E30253 Temperature sensor missing / defective

Code Description

E30301 Standstill not possible (timeout after 10 sec)

E30302 Weight value out of range (overload or underload)

E30303 Incorrect calibration weight value

E30304 Negative initial weight

E30305 Wrong initial weight

E30306 ACT_WGT 65535

E30307 Alarm limits permanently violated

E30308 INF_WGT invalid

E30350 Calibration factor missing or incorrect

E30351 Temp. coeff. factors missing or incorrect

E30352 Double fuse defective (zero value)

E30353 Temperature sensor missing / defective

2-34 Fresenius Medical Care multiFiltrate TM 6/03.07

Scales nos. 3 and 4 (both)

Chapter 2: Functional Description

Code Description

E30401 Standstill not possible (timeout after 10 sec)

E30402 Weight value out of range (overload or underload)

E30403 Incorrect calibration weight value

E30404 Negative initial weight

E30405 Wrong initial weight

E30406 ACT_WGT 65535

E30407 Alarm limits permanently violated

E30408 INF_WGT invalid

E30450 Calibration factor missing or incorrect

E30451 Temp. coeff. factors missing or incorrect

E30452 Double fuse defective (zero value)

E30453 Temperature sensor missing / defective

2.5.5 Error Codes of Pressures Lo-Level Routines

PART

Code Description

E30551 Calibration factor missing, incorrect

E30552 Offset factor missing, incorrect

E30553 Offset > measurement value

E30557 DAC setting missing, incorrect

PPHF

Code Description

E30651 Calibration factor missing, incorrect

E30652 Offset factor missing, incorrect

E30653 Offset > measurement value

E30657 DAC setting missing, incorrect

Fresenius Medical Care multiFiltrate TM 6/03.07 2-35

Loading…

Наборы расходных материалов для аппарата multiFiltrate

Наборы расходных материалов для аппарата multiFiltrate содержат:

  • multiFiltrate — кассету с интегрированными магистралями крови и фильтрата или комплект магистралей с уменьшенным экстракорпоральным объёмом или педиатрический набор магистралей с минимальным объёмом экстракорпорального контура
  • Магистраль замещающего и/или диализирующего растворов (в зависимости от типа процедуры)
  • Гемофильттр или плазмафильтр
     

multiFiltrate — кассета с артериальной, венозной магистралью, магистралью фильтрата, скомпонованными на единой платформе, позволяет с легкостью установить экстракорпоральный контур для каждой процедуры. Комбинация multiFiltrate — кассеты с магистралями, обеспечивающими проведение определенного типа терапии, гарантирует быструю и безопасную подготовку аппарата к работе.

Your Donation Will Be Matched 1-to-1! Can You Chip In?

Dear Patron: Please don’t scroll past this. The Internet Archive is a nonprofit fighting for universal access to quality information. We build and maintain all our own systems, but we don’t charge for access, sell user information, or run ads. Instead, we’re powered by online donations averaging about $14. We’d be deeply grateful if you’d join the one in a thousand users that support us financially.

Right now, we have a matching gift campaign that will double the impact of every donation. We understand that not everyone can donate right now, but if you can afford to contribute this Monday, we promise it will be put to good use. Our resources are crucial for knowledge lovers everywhere—so if you find all these bits and bytes useful, please pitch in.

Your Donation Will Be Matched! Can You Chip In?
Dear Patron: Please don’t scroll past this. Right now we have a matching gift campaign that will double the impact of every donation. We understand that not everyone can give right now, but if you can afford to contribute this Monday, we promise it will be put to good use. If you find all these bits and bytes useful, please pitch in.

Download or browse on-line these User Manual Manual for Fresenius Medical Care multiFiltrate Medical Equipment.

Summary of Contents:

[Page 1] Fresenius Medical Care multiFiltrate

multiFiltrate Instructions for Use Software version: 5.3 Edition: 15A-2015 Part no.: M56 963 1

[Page 2] Fresenius Medical Care multiFiltrate

[Page 3] Fresenius Medical Care multiFiltrate

Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 iii Table of contents 1Index 2 Important information 2.1 How to use the Instructions for Use……………………………………………………………………………. 2-1 2.2 Sign…

[Page 4] Fresenius Medical Care multiFiltrate

iv Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 3.2.1 Monitor………………………………………………………………………………………………………………………. 3-4 3.2.2 Heparin pump………………….

[Page 5] Fresenius Medical Care multiFiltrate

Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 v 4.5.7.3 Treatment parameters…………………………………………………………………………………… 4-27 4.5.8 Change of procedure (except for CVVHD and CVVH) ………..

[Page 6] Fresenius Medical Care multiFiltrate

vi Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4.7.7.1 Treatment main screen …………………………………………………………………………………. 4-59 4.7.7.2 Treatment menu …………………………………..

[Page 7] Fresenius Medical Care multiFiltrate

Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 vii 4.9.7.1 Terminating the treatment ……………………………………………………………………………… 4-89 4.9.7.2 Starting reinfusion ……………………………..

[Page 8] Fresenius Medical Care multiFiltrate

viii Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 5 Alarm processing 5.1 Acknowledging a message repeatedly………………………………………………………………………… 5-1 5.2 Alarm schemes ……………………….

[Page 9] Fresenius Medical Care multiFiltrate

Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 ix 7.3.2.3 Control of systemic ionised calcium ………………………………………………………………… 7-23 7.3.2.4 Controlling the systemic acid-base status………………..

[Page 10] Fresenius Medical Care multiFiltrate

x Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 12 Specifications 12.1 Dimensions and weight…………………………………………………………………………………………….. 12-1 12.2 Identification label …………

[Page 11] Fresenius Medical Care multiFiltrate

Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 xi 15.2 Instructions on the use of «free software»…………………………………………………………………. 15-2

[Page 12] Fresenius Medical Care multiFiltrate

xii Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015

[Page 13] Fresenius Medical Care multiFiltrate

Chapter 1: Index Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 1-1 1 Index A Abbreviations 13-3 Accessories 8-1, 8-6 Acute haemodialysis treatment 13-1 Additional equipment 8-6 Additional information 5-7 Addresses 2-16 Air detector …

[Page 14] Fresenius Medical Care multiFiltrate

Chapter 1: Index 1-2 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 E Earthing contact 3-8 Effluent 7-9 Electrical safety 12-3 Electrical supply 12-4 Electromagnetic compatibility 12-5 Electromagnetic interference immunity 12-7 Electr…

[Page 15] Fresenius Medical Care multiFiltrate

Chapter 1: Index Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 1-3 Previous treatment, continuing 4-15 Procedure, changing 4-28, 4-29 Protection class 12-3 Protective earth conductor 9-3 R Radiated interference source 12-5 Ratio of cal…

[Page 16] Fresenius Medical Care multiFiltrate

Chapter 1: Index 1-4 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015

[Page 17] Fresenius Medical Care multiFiltrate

Chapter 2: Important information Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 2-1 2 Important information 2.1 How to use the Instructions for Use Device designation The multiFiltrate will be identified with «device» hereinafte…

[Page 18] Fresenius Medical Care multiFiltrate

Chapter 2: Important information 2-2 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 Before the responsible organisation can begin to operate the device, the individual responsible for the operation must have been instructed by the manufac…

[Page 19] Fresenius Medical Care multiFiltrate

Chapter 2: Important information Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 2-3 2.5 Brief description The device enables the performance of extracorporeal blood purification without additional equipment. Depending on the procedure use…

[Page 20] Fresenius Medical Care multiFiltrate

Chapter 2: Important information 2-4 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 2.6.3 Treatment modes and fields of application SCUF – Diuretic resistant fluid retention, especially in the event of cardiac insufficiency or pulmonary …

[Page 21] Fresenius Medical Care multiFiltrate

Chapter 2: Important information Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 2-5 2.6.5 Paediatric CVVH and CVVHD treatment modes The dose of the renal replacement therapy for a paediatric treatment can be derived from the recommendation…

[Page 22] Fresenius Medical Care multiFiltrate

Chapter 2: Important information 2-6 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 For more detailed information on situations where these risks might occur and on the options of reducing the occurrence of these risks, please refer to (s…

[Page 23] Fresenius Medical Care multiFiltrate

Chapter 2: Important information Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 2-7 2.6.8 Interaction with other systems The use of tube roller pumps may lead to minimal electrostatic discharge into the tubing system due to friction on the…

[Page 24] Fresenius Medical Care multiFiltrate

Chapter 2: Important information 2-8 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 2.7 To be observed when working on the device Further information on installation (see chapter 9 on page 9-1). Further information on the Technical Safety C…

[Page 25] Fresenius Medical Care multiFiltrate

Chapter 2: Important information Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 2-9 The manufacturer offers training on this device. The local service support organisation is available if you have further questions (see chapter 2.13 on p…

[Page 26] Fresenius Medical Care multiFiltrate

Chapter 2: Important information 2-10 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 2.11 Disclaimer of liability 2.12 Warnings 2.12.1 Warnings relating to hygiene 2.12.2 Warnings related to therapy Warning The device has been approved for…

[Page 27] Fresenius Medical Care multiFiltrate

Chapter 2: Important information Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 2-11 Warning Risk of cross-contamination in the event of an incorrect response to a wet or defective hydrophobic filter  Never force back any fluid with a s…

[Page 28] Fresenius Medical Care multiFiltrate

Chapter 2: Important information 2-12 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 Warning During the treatment, check the following at appropriate intervals: – The condition of the patient. – The function of the device and the extr…

[Page 29] Fresenius Medical Care multiFiltrate

Chapter 2: Important information Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 2-13 2.12.3 Warnings relating to the system Warning Risk of injury due to device defect If the device has the following defects, the measures given below must…

[Page 30] Fresenius Medical Care multiFiltrate

Chapter 2: Important information 2-14 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 2.12.4 Electrical warnings The touch current or the patient leakage current from another non-medical electrical device or medical electrical device in the…

[Page 31] Fresenius Medical Care multiFiltrate

Chapter 2: Important information Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 2-15 The alarm limits of the patient leakage current for type CF applied parts are as follows: – 10 µA AC / DC (normal case, error-free) – 50 µA AC / DC …

[Page 32] Fresenius Medical Care multiFiltrate

Chapter 2: Important information 2-16 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 2.13 Addresses Manufacturer Fresenius Medical Care AG & Co. KGaA 61346 Bad Homburg Germany Phone: +49 6172 609-0 www.fmc-ag.com International service s…

[Page 33] Fresenius Medical Care multiFiltrate

Chapter 3: Design Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 3-1 3Design 3.1 Views of the device 3.1.1 Front view Legend 1 IV pole 2 Scale 1 3 Scale 2 4 Monitor 5 Extracorporeal blood circuit module 6 Scales 3 and 4 4 5 6 1 3 2

[Page 34] Fresenius Medical Care multiFiltrate

Chapter 3: Design 3-2 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 3.1.2 Rear view Legend 1 Tube holders 2 Heparin pump 3 Identification label 4 Connector strip 5 Pocket for Instructions for Use 3 4 1 2 5

[Page 35] Fresenius Medical Care multiFiltrate

Chapter 3: Design Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 3-3 3.1.2.1 Connector strip Legend 1 Power switch 2 Line voltage selector 3 Power supply connection 4 Equipotential bonding 5 10-Base-T Ethernet socket (LAN) (option) 6 RS 232…

[Page 36] Fresenius Medical Care multiFiltrate

Chapter 3: Design 3-4 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 3.2 Controls and indicators 3.2.1 Monitor Legend 1 Screen 2 Tube holders 3 Rotary selector and [OK] key 4 [ESC] key 5 [Mute] key 6[STOP] key 7 [START/RESET] key 8[On/Off] …

[Page 37] Fresenius Medical Care multiFiltrate

Chapter 3: Design Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 3-5 3.2.2 Heparin pump Legend 1 The  key moves the slide carriage up as long as the key is pressed. 2 The  key moves the slide carriage automatically down. Pressing th…

[Page 38] Fresenius Medical Care multiFiltrate

Chapter 3: Design 3-6 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 3.2.3 Extracorporeal blood circuit module Legend 1 Dialysate heater (green dot) 2 Dialysate pump (green dot) 3 Tube holder 4 Substituate pump (white dot) 5 Substituate hea…

[Page 39] Fresenius Medical Care multiFiltrate

Chapter 3: Design Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 3-7 3.2.4 multiFiltrate with Ci-Ca module (option) Legend 1 Second IV pole with connector in the IV pole bearing 2 Ci-Ca module 2 1

[Page 40] Fresenius Medical Care multiFiltrate

Chapter 3: Design 3-8 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 3.2.4.1 Ci-Ca module front view Legend 1 Citrate drop counter 2 Citrate pump 3 Citrate insertion switch 4 Calcium drop counter 5 Calcium pump 6 Calcium insertion switch 3…

[Page 41] Fresenius Medical Care multiFiltrate

Chapter 3: Design Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 3-9 3.3 User interface Legend A Status bar Can have the following colours: Blue: in preparation Green: during treatment Magenta: end of treatment Grey: system parameters and s…

[Page 42] Fresenius Medical Care multiFiltrate

Chapter 3: Design 3-10 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015

[Page 43] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-1 4Operation 4.1 Application principles  Applicable to devices with 10-Base-T Ethernet connection (LAN)  Applicable to all devices With regard to the application pr…

[Page 44] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-2 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 Warning Risk of cross-contamination if tubing systems without hydrophobic filters are used  Only use tubing systems with hydrophobic filters in the pressure lines…

[Page 45] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-3 Warning When inserting the tubing systems, observe the following: The absence of air, kinks, tensions, and twists in the tubing system and the correct position of al…

[Page 46] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-4 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 Warning During the treatment, check the following at appropriate intervals: – The condition of the patient. – The function of the device and the extracorporeal bl…

[Page 47] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-5 4.2 Basic operation philosophy 4.2.1 Menu selection Pressing the [ESC] key exits the current menu field and activates the menu bar. The selected menu has a green backg…

[Page 48] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-6 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 A new menu field (green background) can be selected by turning the rotary selector [OK]. Turning the rotary selector to the right will select the menu fields on the r…

[Page 49] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-7 4.2.3 Entering treatment parameters A parameter field (green background) can be selected by turning the rotary selector. Turning the rotary selector to the right will…

[Page 50] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-8 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 The target value is set, for example, to 200 ml/min by turning the rotary selector (the target value is shown in white). The selected value must be confirmed by pres…

[Page 51] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-9 4.3 Basic operation steps 4.3.1 Entering ultrafiltration / UF goal Note The parameter UF goal can only be entered in the Treatment parameters menu. If ultrafiltration…

[Page 52] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-10 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4.3.2 Resetting the alarm limits in alarm-free condition The UF goal field is used to configure the currently planned target goal. Under «Total», the system…

[Page 53] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-11 4.3.3 Raising and lowering the venous bubble catcher level 4.3.3.1 Raising the level 4.3.3.2 Lowering the level 4.3.4 Bolus anticoagulation The bolus anticoagulation f…

[Page 54] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-12 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4.4 Making the device ready for operation 4.4.1 Turning the device on  Use the rotary selector to select Bolus and press and hold [OK]. As long as the selector i…

[Page 55] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-13 Note After turning the device off, wait for approximately 10 seconds before turning it on again.  Connect the device to the power supply.  Connect the equipote…

[Page 56] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-14 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4.4.2 Starting the functional test  Use the rotary selector to select All conditions fulfilled? [OK] to confirm! and press [OK]. Note If the multiFiltrate device…

[Page 57] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-15 4.4.3 Selecting the anticoagulation method 4.4.4 Continuing the previous treatment Note The following screen will not appear if: – The Ci-Ca module is not attached….

[Page 58] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-16 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4.4.5 Selecting the treatment mode When the treatment is continued with the previous treatment parameters, the operator has the option to start with or without using…

[Page 59] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-17 CRRT treatments 4.5 CRRT treatments General description of the CVVH, CVVHD, CVVHDF, and Pre-Post CVVH treatments with information on the differences between the indi…

[Page 60] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-18 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 CRRT treatments 4.5.2.2 Inserting the arterial blood line system 4.5.2.3 Inserting the venous blood line system  Insert the arterial blood line system according to…

[Page 61] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-19 CRRT treatments 4.5.2.4 Inserting the filtrate line system 4.5.3 Inserting further tubing systems Depending on the selected treatment mode, the tubing system for dial…

[Page 62] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-20 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 CRRT treatments 4.5.3.1 Inserting the dialysate or sub predilution tubing system (except for CVVH) 4.5.3.2 Inserting the substituate or sub postdilution tubing system …

[Page 63] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-21 CRRT treatments 4.5.4 Inserting the heparin syringe 4.5.5 Complete tubing arrangement Note If a heparin syringe is used, it must be inserted and connected before con…

[Page 64] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-22 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 CRRT treatments 4.5.6 Preparation 4.5.6.1 Priming the tubing system This screen will be displayed immediately if the setting in the System parameters menu was set to…

[Page 65] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-23 CRRT treatments 4.5.6.2 Rinsing the tubing system / entering treatment parameters  Open the infusion / extraction point to prime the arterial bubble catcher. If th…

[Page 66] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-24 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 CRRT treatments 4.5.6.3 UF rinse 4.5.6.4 Recirculation / waiting for patient Note When using NaCl solutions with only one connector, make sure there is enough NaCl …

[Page 67] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-25 CRRT treatments 4.5.6.5 Connecting the patient Tip If the patient is not yet available, then recirculation can be continued by pressing the [START/RESET] key. Note …

[Page 68] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-26 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 CRRT treatments 4.5.7 Treatment 4.5.7.1 Treatment main screen The optical detector senses opaque fluid. The blood pump is stopped. If not done yet when using an NaCl …

[Page 69] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-27 CRRT treatments 4.5.7.2 Treatment menu 4.5.7.3 Treatment parameters  Press the [ESC] key.  Use the rotary selector to select Treatment from the menu bar and pre…

[Page 70] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-28 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 CRRT treatments 4.5.8 Change of procedure (except for CVVHD and CVVH) 4.5.8.1 Change of procedure to CVVH 4.5.8.2 Change of procedure to CVVHD Note It is possible to …

[Page 71] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-29 CRRT treatments 4.5.9 Change of procedure back to CVVHDF  Use the rotary selector to select Substituate and set it to Off. Warning The storage life for opened sol…

[Page 72] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-30 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 CRRT treatments 4.5.10 End of treatment 4.5.10.1 Terminating the treatment 4.5.10.2 Starting reinfusion  Press the [ESC] key.  Use the rotary selector to select …

[Page 73] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-31 CRRT treatments Indication of the decreasing reinfusion volume. Selecting View treatment history? [OK] to confirm! will display the treatment parameters. The optica…

[Page 74] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-32 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 CRRT treatments 4.5.11 Disconnecting the patient and removing the tubing system 4.5.12 Treatment history Warning Consumables must be discarded after the treatment in…

[Page 75] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-33 CRRT treatments with citrate anticoagulation 4.6 CRRT treatments with citrate anticoagulation General description of the Ci-Ca CVVHD and Ci-Ca postCVVHDF treatments w…

[Page 76] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-34 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 CRRT treatments with citrate anticoagulation 4.6.2 Inserting the cassette system 4.6.2.1 Inserting the citrate tube and the calcium tube  Insert the cassette system…

[Page 77] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-35 CRRT treatments with citrate anticoagulation 4.6.3 Inserting further tubing systems Depending on the selected treatment mode, the tubing system for dialysate and / or…

[Page 78] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-36 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 CRRT treatments with citrate anticoagulation 4.6.3.2 Inserting the substituate tubing system (except for CVVHD Ci-Ca)  Insert the dialysate tubing system according…

[Page 79] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-37 CRRT treatments with citrate anticoagulation 4.6.4 Inserting the heparin syringe 4.6.5 Complete tubing arrangement Note If a heparin syringe is used, it must be inser…

[Page 80] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-38 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 CRRT treatments with citrate anticoagulation 4.6.6 Preparation 4.6.6.1 Filling the citrate and calcium drip chambers This screen will be displayed immediately if the …

[Page 81] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-39 CRRT treatments with citrate anticoagulation 4.6.6.2 Priming the Ci-Ca lines The Ci-Ca pumps fill the Ci-Ca lines with fluid for a defined time period. In that time p…

[Page 82] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-40 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 CRRT treatments with citrate anticoagulation 4.6.6.3 Priming the tubing system 4.6.6.4 Rinsing the tubing system / entering treatment parameters  Use the rotary sel…

[Page 83] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-41 CRRT treatments with citrate anticoagulation Note The heparin dose is to be administered as prescribed by the physician! If an initial heparin dose is to be administe…

[Page 84] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-42 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 CRRT treatments with citrate anticoagulation 4.6.6.5 UF rinse 4.6.6.6 Recirculation / waiting for patient Note When using NaCl solutions with only one connector, mak…

[Page 85] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-43 CRRT treatments with citrate anticoagulation 4.6.6.7 Connecting the patient Tip If the patient is not yet available, then recirculation can be continued by pressing …

[Page 86] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-44 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 CRRT treatments with citrate anticoagulation 4.6.7 Treatment 4.6.7.1 Checking the post-filter calcium concentration The optical detector senses opaque fluid. The blood…

[Page 87] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-45 CRRT treatments with citrate anticoagulation 4.6.7.2 Treatment main screen 5 minutes after the treatment has been started, a note is displayed, accompanied by a shor…

[Page 88] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-46 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 CRRT treatments with citrate anticoagulation 4.6.7.3 Treatment menu 4.6.7.4 Treatment parameters  Press the [ESC] key.  Use the rotary selector to select Treatm…

[Page 89] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-47 CRRT treatments with citrate anticoagulation 4.6.8 End of treatment 4.6.8.1 Terminating the treatment 4.6.8.2 Starting reinfusion  Press the [ESC] key.  Use the …

[Page 90] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-48 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 CRRT treatments with citrate anticoagulation 4.6.9 Disconnecting the patient and removing the tubing system Indication of the decreasing reinfusion volume. Selecting …

[Page 91] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-49 CRRT treatments with citrate anticoagulation  Disconnect the venous patient line from the patient.  Remove the calcium pump segment from the stationary part (s…

[Page 92] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-50 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 CRRT treatments with citrate anticoagulation 4.6.10 Treatment history Indication of the treatment parameters for the entire treatment.  Press the [I/O] key to turn…

[Page 93] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-51 Paediatric CRRT treatments 4.7 Paediatric CRRT treatments General description of the paediatric CVVH and CVVHD treatments with information on the differences between …

[Page 94] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-52 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 Paediatric CRRT treatments 4.7.2.2 Inserting the venous blood line system 4.7.2.3 Inserting the filtrate line system 4.7.3 Inserting further tubing systems Depending o…

[Page 95] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-53 Paediatric CRRT treatments 4.7.3.1 Inserting the dialysate tubing system (except for CVVH) 4.7.3.2 Inserting the substituate tubing system (except for CVVHD) Note Dep…

[Page 96] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-54 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 Paediatric CRRT treatments 4.7.4 Inserting the heparin syringe 4.7.5 Complete tubing arrangement Note If a heparin syringe is used, it must be inserted and connected …

[Page 97] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-55 Paediatric CRRT treatments 4.7.6 Preparation 4.7.6.1 Priming the tubing system This screen will be displayed immediately if the setting in the System parameters menu …

[Page 98] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-56 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 Paediatric CRRT treatments 4.7.6.2 Rinsing the tubing system / entering treatment parameters Note The heparin dose is to be administered as prescribed by the physicia…

[Page 99] Fresenius Medical Care multiFiltrate

Chapter 4: Operation Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-57 Paediatric CRRT treatments 4.7.6.3 UF rinse 4.7.6.4 Recirculation / waiting for patient Note When using NaCl solutions with only one connector, make sure there is e…

[Page 100] Fresenius Medical Care multiFiltrate

Chapter 4: Operation 4-58 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 Paediatric CRRT treatments 4.7.6.5 Connecting the patient Tip If the patient is not yet available, then recirculation can be continued by pressing the [START/RESET] …

multiFiltrate

Service Manual

Technical Manual

Edition: 6/03.07

Part no. M28 003 1

Table of Contents

1 Important Information

1.1 Organization of the Technical Manual …………………………………………………………………………. 1-1

1.2 How to Use the Technical Manual……………………………………………………………………………….. 1-2

1.3 Precautions for Working on the System ……………………………………………………………………… 1-3

1.4 Addresses …………………………………………………………………………………………………………………. 1-4

2 Functional Description

2.1 Extracorporeal Circuit………………………………………………………………………………………………… 2-1

2.1.1 Pumps ……………………………………………………………………………………………………………………….. 2-1

2.1.2 Heaters ………………………………………………………………………………………………………………………. 2-2

2.1.3 Pressure Transducer ……………………………………………………………………………………………………. 2-2

2.1.4 Air Detector and Venous Clamp …………………………………………………………………………………….. 2-2

2.1.5 Optical Detector (Non-Opaque/Opaque Fluid Detector)…………………………………………………….. 2-3

2.1.6 Blood Leak Detector …………………………………………………………………………………………………….. 2-3

2.1.7 Heparin Pump……………………………………………………………………………………………………………… 2-3

2.2 Weighing Units ………………………………………………………………………………………………………….. 2-3

2.3 Ci-Ca Module (Option)………………………………………………………………………………………………… 2-4

2.4 Functional Test (T1 Test) and Error Messages…………………………………………………………….. 2-4

2.4.1 Battery Test, Part 1………………………………………………………………………………………………………. 2-5

2.4.2 Scales Test …………………………………………………………………………………………………………………. 2-5

2.4.3 Pump Test ………………………………………………………………………………………………………………….. 2-6

2.4.4 Pressure Transducer ……………………………………………………………………………………………………. 2-8

2.4.5 Optical Detector…………………………………………………………………………………………………………… 2-9

2.4.6 Air Detector……………………………………………………………………………………………………………….. 2-10

2.4.7 Blood Leak Detector …………………………………………………………………………………………………… 2-11

2.4.8 Heater ………………………………………………………………………………………………………………………. 2-12

2.4.9 Battery Test, Part 2…………………………………………………………………………………………………….. 2-15

2.4.10 Audible Alarm ……………………………………………………………………………………………………………. 2-16

2.4.11 Heparin Pump ……………………………………………………………………………………………………………. 2-17

2.4.12 multiDataLink…………………………………………………………………………………………………………….. 2-18

2.4.13 Ci-Ca Module (Option)………………………………………………………………………………………………… 2-18

2.5 Error Messages………………………………………………………………………………………………………… 2-20

2.5.1 Alarm Messages………………………………………………………………………………………………………… 2-20

2.5.2 Warning Messages…………………………………………………………………………………………………….. 2-22

2.5.3 Fatal Errors ……………………………………………………………………………………………………………….. 2-30

2.5.4 Error Codes of Scales Lo-Level Routines ……………………………………………………………………. 2-33

2.5.5 Error Codes of Pressures Lo-Level Routines……………………………………………………………….. 2-35

2.5.6 Error Codes of Heparin Pump Lo-Level Routines…………………………………………………………. 2-36

2.5.7 Error Codes of Pumps Lo-Level Routines……………………………………………………………………. 2-37

2.6 Overview of Display Identification Numbers ……………………………………………………………… 2-37

Fresenius Medical Care multiFiltrate TM 6/03.07 0-1

3 Installation

3.1 Preface………………………………………………………………………………………………………………………. 3-1

3.2 Important Information on Initial Start-Up …………………………………………………………………….. 3-2

3.3 Initial Start-Up Report multiFiltrate……………………………………………………………………………… 3-3

3.4 Explanations on the Initial Start-Up Report …………………………………………………………………. 3-8

3.5 Installing the Ci-Ca Module (Option)………………………………………………………………………….. 3-16

4 TSC / TMC / Maintenance

4.1 Important Information Regarding the Procedure …………………………………………………………. 4-1

4.2 TSC / MA Report multiFiltrate……………………………………………………………………………………… 4-2

4.3 multiFiltrate TSC Report …………………………………………………………………………………………….. 4-8

4.4 Explanations on the TSC / MA Report ……………………………………………………………………….. 4-12

5 Adjustment Instructions and Tests

5.1 Service Tools …………………………………………………………………………………………………………….. 5-1

5.2 Service Program ………………………………………………………………………………………………………… 5-2

5.2.1 Start …………………………………………………………………………………………………………………………… 5-2

5.2.2 Selecting the Language ………………………………………………………………………………………………… 5-3

5.2.3 System Messages ……………………………………………………………………………………………………….. 5-3

5.2.4 Deactivating and Activating the Heparin Pump ………………………………………………………………… 5-3

5.2.5 Filtrate Bag Monitoring Limit ………………………………………………………………………………………….. 5-4

5.2.6 Dialysate Tubing Arrangement ………………………………………………………………………………………. 5-4

5.2.7 Option MultiDataLink (mDL) ………………………………………………………………………………………….. 5-5

5.2.8 Taring and Calibrating the Scales…………………………………………………………………………………… 5-7

5.2.9 Calibrating the Pressures ……………………………………………………………………………………………… 5-8

5.2.10 Calibrating the Blood Leak Detector ……………………………………………………………………………… 5-10

5.2.11 Verifying the System Values………………………………………………………………………………………… 5-11

5.2.12 Events Memory………………………………………………………………………………………………………….. 5-13

5.2.13 Program Treatment Modes………………………………………………………………………………………….. 5-15

5.2.14 Programming the Ci-Ca Data (Option) ………………………………………………………………………….. 5-16

5.3 Extracorporeal Components …………………………………………………………………………………….. 5-17

5.3.1 Tightness of the Venous Occlusion Clamp…………………………………………………………………….. 5-17

5.3.2 Checking the Pressure Transducers …………………………………………………………………………….. 5-17

5.3.3 Venous Pressure Transducer (P.C.B. LP 450-3)…………………………………………………………….. 5-18

5.3.4 Air Detector (P.C.B. LP 450-3) …………………………………………………………………………………….. 5-20

5.3.5 Optical Detector Sensing Opaque / Non-Opaque Fluid (P.C.B. LP 450-3) …………………………. 5-21

5.3.6 Arterial / Filtrate / PHF Pressure Transducers (P.C.B. LP 343-1) ……………………………………… 5-21

5.3.7 Blood Leak Detector …………………………………………………………………………………………………… 5-22

5.3.8 Setting the Heparin Pump……………………………………………………………………………………………. 5-23

5.4 Programming the Processors …………………………………………………………………………………… 5-25

5.4.1 Display Processor (DP) ………………………………………………………………………………………………. 5-26

5.4.2 Operating and Safety Processors (OP and SP) ……………………………………………………………… 5-26

5.4.3 Interchangeability of Safety and Operating Processor …………………………………………………….. 5-30

5.5 Ci-Ca Module (Option) ………………………………………………………………………………………………. 5-31

0-2 Fresenius Medical Care multiFiltrate TM 6/03.07

5.5.1 Preparing the Functional Test ……………………………………………………………………………………… 5-31

5.5.2 Functional Test of the Drip Counter………………………………………………………………………………. 5-32

5.5.3 Functional Test of the Hall Sensor ……………………………………………………………………………….. 5-32

5.5.4 Functional Test of the Insertion Switch………………………………………………………………………….. 5-32

5.5.5 Functional Test of the Line Occlusion …………………………………………………………………………… 5-33

6 PC Service Software MFT

6.1 Organization of the Quick Guide…………………………………………………………………………………. 6-1

6.2 General Information …………………………………………………………………………………………………… 6-1

6.3 Preparation………………………………………………………………………………………………………………… 6-1

6.3.1 System Requirements ………………………………………………………………………………………………….. 6-1

6.3.2 Software Installation …………………………………………………………………………………………………….. 6-2

6.3.3 Hardware Installation……………………………………………………………………………………………………. 6-3

6.3.4 ServiceCard Description ……………………………………………………………………………………………….. 6-3

6.3.5 Starting the Software ……………………………………………………………………………………………………. 6-3

7 Block Diagrams and Component Layouts

7.1 Block Diagram …………………………………………………………………………………………………………… 7-2

7.2 AC Wiring ………………………………………………………………………………………………………………….. 7-3

7.3 P.C.B. LP 122 Heater Control ……………………………………………………………………………………… 7-4

7.3.1 Block Diagram …………………………………………………………………………………………………………….. 7-4

7.3.2 Description………………………………………………………………………………………………………………….. 7-4

7.3.3 Component Layout ………………………………………………………………………………………………………. 7-5

7.4 P.C.B. LP 123 Pump Control……………………………………………………………………………………….. 7-6

7.4.1 Description………………………………………………………………………………………………………………….. 7-6

7.4.2 Component Layout ………………………………………………………………………………………………………. 7-7

7.5 P.C.B. LP 124 Motherboard ………………………………………………………………………………………… 7-8

7.5.1 Description………………………………………………………………………………………………………………….. 7-8

7.5.2 Component Layout ………………………………………………………………………………………………………. 7-9

7.6 P.C.B. LP 125 Blood Leak Detector …………………………………………………………………………… 7-10

7.6.1 Component Layout …………………………………………………………………………………………………….. 7-10

7.7 P.C.B. LP 127 Scales Board………………………………………………………………………………………. 7-11

7.7.1 Description………………………………………………………………………………………………………………… 7-11

7.7.2 Component Layout …………………………………………………………………………………………………….. 7-11

7.8 P.C.B. LP 128, Power Supply Unit……………………………………………………………………………… 7-12

7.8.1 Block Diagram …………………………………………………………………………………………………………… 7-12

7.8.2 Description………………………………………………………………………………………………………………… 7-12

7.8.3 Component Layout …………………………………………………………………………………………………….. 7-13

7.9 P.C.B. LP 129 User Interface …………………………………………………………………………………….. 7-14

7.9.1 Component Layout …………………………………………………………………………………………………….. 7-14

7.10 P.C.B. LP 244 Operating and Safety Processors………………………………………………………… 7-15

7.10.1 Block Diagram …………………………………………………………………………………………………………… 7-15

7.10.2 Jumper Description…………………………………………………………………………………………………….. 7-15

7.10.3 Component Layout …………………………………………………………………………………………………….. 7-17

Fresenius Medical Care multiFiltrate TM 6/03.07 0-3

7.11 P.C.B LP 343-4 Pressure Transducer ………………………………………………………………………… 7-18

7.11.1 Component Layout …………………………………………………………………………………………………….. 7-18

7.12 P.C.B. LP 450-3 multiFiltrate Air Detector Control ……………………………………………………… 7-19

7.12.1 Component Layout …………………………………………………………………………………………………….. 7-19

7.13 P.C.B. LP 950, Control Board (Heparin Pump)……………………………………………………………. 7-20

7.13.1 Component Layout …………………………………………………………………………………………………….. 7-20

7.14 Ci-Ca Module……………………………………………………………………………………………………………. 7-21

7.14.1 Block Diagram …………………………………………………………………………………………………………… 7-21

7.14.2 Component Layout (P.C.B. LP MS 0407) ………………………………………………………………………. 7-22

0-4 Fresenius Medical Care multiFiltrate TM 6/03.07

1 Important Information

1.1 Organization of the Technical Manual

Page identification Page number 1-3 is to be interpreted as: Chapter 1, page 3.

Editorial information The current edition of this Technical Manual is

6/03.07 = 5th edition, September 2006

In case of updates, the chapters concerned will be replaced.

Refer to the table below to verify that the Technical Manual is up-to­date.

Chapter Current version

1 6/03.07

2 6/03.07

3 6/03.07

4 6/03.07

5 6/03.07

6 6/03.07

7 6/03.07

Chapter 1: Important Information

Changes Manual changes will be released as new editions and supplements. In

general — subject to change without notice.

Fresenius Medical Care multiFiltrate TM 6/03.07 1-1

Chapter 1: Important Information

1.2 How to Use the Technical Manual

Purpose This Technical Manual is intended for service technicians and is to be

used for first studies (to acquire a basic knowledge) and for reference
purposes (for TSC, maintenance and repair). The Technical Manual,
however, does not replace the training courses offered by the
manufacturer.

Requirements Knowledge of the current Operating Instructions for the respective

system.
Background experience in mechanics, electrical and medical
engineering.

Specifications For the specifications of the respective system, refer to the current

Operating Instructions.

Circuit diagrams and
component layouts

Explanation of the Note
and Caution symbols used

The identification on the PCB permits the operator/technician to verify if
the circuit diagram/component layout matches the PCB actually
installed in the system.

Note

Informs the operator that in case of a failure to follow the steps as
described, a specific function will be executed incorrectly or will not be
executed at all, or will not produce the desired effect.

Caution

Advises the operator against certain procedures or actions that could
cause damage to the equipment or may have adverse effects on
operators and patients.

1-2 Fresenius Medical Care multiFiltrate TM 6/03.07

1.3 Precautions for Working on the System

Authorized persons Assembly, extensions, adjustments, modifications or repairs may only

be carried out by the manufacturer or persons authorized by him.

Chapter 1: Important Information

Test equipment and
accessories

Precautions When working on the open system, the following precautions must be

ESD precautions When repairing and when replacing spare parts, observe the applicable

The activities described in the Technical Manual require the availability
of the necessary technical test equipment and accessories.

respected:
 Protect the components against ingress of fluids.
 Do not touch live parts.
 All plugs, connections and components may only be disconnected or

connected if de-energized.

ESD precautions (e.g. EN 100 015-1).

Fresenius Medical Care multiFiltrate TM 6/03.07 1-3

Chapter 1: Important Information

1.4 Addresses

Manufacturer Fresenius Medical Care AG & Co. KGaA

Please address any inquiries to:

D-61346 Bad Homburg
Germany
+49 (0)6172/609-0
www.fmc-ag.com

Service
Central Europe

International
service

Local service

Fresenius Medical Care
Deutschland GmbH
Geschäftsbereich Zentraleuropa
Kundendienst / Servicecenter
Steinmühlstrasse 24b
61352 Bad Homburg
Germany
Phone: +49 6172 609-7100
Fax: +49 6172 609-7102
E-mail: ServicecenterD@fmc-ag.com

Fresenius Medical Care
Deutschland GmbH
Service Support International
Hafenstrasse 9
D-97424 Schweinfurt
Germany
Phone: +49 9721 678-333 (hotline)
Fax: +49 9721 678-130

1-4 Fresenius Medical Care multiFiltrate TM 6/03.07

2 Functional Description

2.1 Extracorporeal Circuit

The elements for maintaining and monitoring the extracorporeal circuit
of the multiFiltrate are as follows:

 Pumps

 Heaters

 Pressure transducer

 Air detector

 Venous clamp

 Non-opaque/opaque fluid detector

 Blood leak detector

 Heparin pump

Chapter 2: Functional Description

2.1.1 Pumps

Altogether, the multiFiltrate is provided with four pumps:

Pump Delivery rate

Blood pump 10  500 ml/min

Filtrate pump 10  180 ml/min

Substituate pump 10  170 ml/min

Dialysate pump 10  170 ml/min

The pumps are driven by direct-current geared motors. To control the
speed, these motors are each provided with a clock pulse generator,
which is directly connected to the motor shaft. The pump processors for
controlling the individual pumps (P.C.B. LP 123) are fitted on the motor
housing.

All pumps are supplied with 24 V. The nominal voltage of the blood
pump motor is 20 V. The nominal voltage of the other pumps, however,
is 24 V.

A Hall sensor in the pump housing and a permanent magnet in each
pump door monitor the state of the pump door.

The line inserting position of the pump rotors for inserting and removing
the pump segments is detected by a combination of a reed switch (in
the pump housing) and a permanent magnet (in the rotor).

The functional test covers a check of all pumps.

Fresenius Medical Care multiFiltrate TM 6/03.07 2-1

Chapter 2: Functional Description

2.1.2 Heaters

2.1.3 Pressure Transducer

To allow heating of replacement fluids, two heaters, which are activated
and monitored independently of each other, are installed in the
multiFiltrate.

The heater foil is applied to the outside of the heater rod and supplied
with approx. 28 V. The temperatures are controlled and monitored by
altogether four NTC sensors (two for the operating processor and two
for the safety processor). This is implemented on P.C.B. LP 122.

Both heaters are tested during the functional test.

Four pressures are measured at the multiFiltrate:

 Arterial pressure,

Measuring transducer (P.C.B. LP 343-1) between the patients
arterial access and the blood pump
Measuring range: 280  300 mmHg

 Pre-hemofilter pressure

Measuring transducer (P.C.B. LP 343-1) between the blood pump
and the filter inlet
Measuring range: 0  750 mmHg

 Venous pressure,

Measuring transducer (P.C.B. LP 450-3) between the filter outlet
and the patients venous access,
Measuring range: 80  500 mmHg.

 Filtrate pressure or dialysate pressure

Measuring transducer (P.C.B. LP 343-1) between the filter
connector and the filtrate pump
Measuring range: 300  300mmHg.

All pressure transducers are subjected to the functional tests.

2.1.4 Air Detector and Venous Clamp

The air detector (P.C.B. LP 450-3) serves for the detection of air in the
extracorporeal blood circuit and operates on the ultrasound principle .
Both the transmitter and the receiver are integrated in the drip chamber
holder. Once the level in the venous drip chamber has fallen below a
certain threshold, the venous clamp is closed. This function is executed
independently of the operating or safety processor. The additional
board AD 28 increases the transmitter voltage during Preparation to
ensure that the level of saline solution is reliably detected.

The air detector is subjected to the functional test.

2-2 Fresenius Medical Care multiFiltrate TM 6/03.07

2.1.5 Optical Detector (Non-Opaque/Opaque Fluid Detector)

The optical detector (P.C.B. LP 450-3) detects, according to the infrared
principle, whether saline solution or blood is present in the tubing
system.

The optical detector is subjected to the functional test.

2.1.6 Blood Leak Detector

The blood leak detector (P.C.B. LP 125) is provided for the detection of
a potential blood loss through the membrane. It is operated applying a
two-color measuring section. In the course of this, red and green light is
alternately transmitted to a reference receiver or, through the filtrate
line, to a measuring receiver.

Both the transmitter and the receiver are integrated in the line holder.

The blood leak detector is subjected to the functional test.

2.1.7 Heparin Pump

Chapter 2: Functional Description

2.2 Weighing Units

The heparin pump is used for continuous heparinization of the blood.

A syringe plunger is moved by means of a carriage bar. The carriage
bar is connected to a threaded spindle via a slide. A microprocessor­controlled stepper motor causes the spindle to rotate. Depending on the
activation, the piston will move up or down.

One Hall sensor each signals when the piston has reached its upper
and lower end of travel. The safety system of the pump comprises a
speed monitoring device (slotted disc with optical sensor) and a motor
current monitoring function. The syringe types are set via a coding
switch (HEX switch).

The weighing units are used for managing the fluid balance during
treatment.

The weighing cells are operated according to the strain gauge principle.
Signal conditioning, including analog-to-digital conversion, is achieved
per weighing cell on the P.C.B. LP 127. The actual weights are
produced by the operating processor.

Scale 1 and scale 2 each have a useful load of approx. 12 kg.

The scales are subjected to the functional test. To test the scales, a test
weight (ball) of defined value must be taken off each scale. Proper
functioning of the scales can be concluded from the correct difference
between the weights before and after lifting.

Fresenius Medical Care multiFiltrate TM 6/03.07 2-3

Chapter 2: Functional Description

2.3 Ci-Ca Module (Option)

The Ci-Ca module is intended for regional citrate anticoagulation in the
CVVHD treatment therapy.

Turning power on The Ci-Ca module requires a supply voltage of 24VDC. This voltage is

provided by the multiFiltrate system’s power supply unit via the
connector in the lower IV pole support of the IV pole located on the right
of the system.
After the multiFiltrate system was turned on by pressing the power
switch on its rear, the supply voltage of the module is connected. The
operating processor of the Ci-Ca module switches into the standby
mode.
If the multiFiltrate systems is then turned on via the I/O key on the front
of the system, the operating processor of the Ci-Ca module will perform
an internal processor test. This test is performed simultaneously to the
processor test of the multiFiltrate system.

The Ci-Ca module communicates with the multiFiltrate basic system via
a serial interface.

Processor test If a test is not passed successfully, the module will not establish

communications with the multiFiltrate system. It is not possible to
perform a treatment with citrate anticoagulation. The multiFiltrate
system recognizes this problem and displays a message which
proposes to use an alternative anticoagulation equipment (e.g. heparin
pump) and which has to be confirmed by the operator.

T1 test The Ci-Ca module performs its own T1 test, independent of the T1 test

of the multiFiltrate system. This test will be started automatically and
simultaneously to the multiFiltrate T1 test after the prompt whether the
starting conditions are met was confirmed with [OK].

This test cannot be skipped or deselected.

System errors If it is still possible, system errors in the Ci-Ca module are shown on the

multiFiltrate display with the indication that citrate anticoagulation is not
available.

2.4 Functional Test (T1 Test) and Error Messages

After it has been turned on the rear and the [I/O] key has been pressed,
the system automatically starts the processor test. After completion of
the processor test, the display test will be performed. In this test, the
numerical characters are represented for 2 seconds in all of the three
fonts used. After this test, the functional test (T1 test) is started
automatically. Depending on the configuration in the SETUP (SETUP
automatic), the test is running in the background or (SETUP detailed)
the test steps are represented separately on the display.

2-4 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

2.4.1 Battery Test, Part 1

At the beginning of the T1 test, the system assumes the terminal voltage
and loads the battery with a defined resistance over the entire test.

2.4.2 Scales Test

In the first step, the temperature compensation and calibration factors
are tested. In the second step, the scales are tared. The linearity test
follows in step three. The 4 test weights are lifted, and the weight (44.3
g) is tested for a tolerance of ±0.5 g. The offset drift is tested in step four.
The test weights are lowered again; the system expects a weight of 0.0
g with a tolerance of ±0.5 g. The time out for each step (2 to 4) is 10 sec.
In the event of an error, messages are emitted. Here, the error
messages section only relates to scales I. The messages are identical
for scales II to IV.

Error message Causes Action required

Scale I

Cal. factor missing, error.
Acknowledge with
[START/RESET] key.

Scale I

Temp. coeff. factors missing, error.
Acknowledge with
[START/RESET] key.

Scale I

Movement detected on Scale I.
Acknowledge with
[START/RESET] key.

Lithium battery on P.C.B. LP 244­OP discharged.

Defect of the EPROM on P.C.B.
LP 127 of the scales specified.

The system is not in a stable
position. The scales are subjected
to a draft. Bags with tubing system
are placed on the scales; this is
especially applicable to filtrate
scales.

The +12 V and/or 12 V voltage(s)
is (are) missing.

Scales mechanics not smooth. Check for smoothness and adjust,

Replace the battery and execute
the calibration step in the Service
program.

Replace the P.C.B. and the
weighing bar and execute the
calibration step in the Service
program.

Set up the system in a stable
position. Remove the system from
the draft, or close the window.
Relieve the scales.

Check the fuses on the PSU board
LP 128 and replace the P.C.B.
LP 128, if necessary.

Caution: All pressures must be
checked and calibrated, if
necessary.

if necessary. Weight check in the
Service program.

Connector not fitted or defective
ribbon cable.

P.C.B. LP 244-OP defective. Replace P.C.B. LP 244-OP.

Fresenius Medical Care multiFiltrate TM 6/03.07 2-5

Fit the connector to the
motherboard and/or to the P.C.B.
LP 127. Replace the ribbon cable.

Caution: All scales and pressures
must be calibrated.

Chapter 2: Functional Description

Error message Causes Action required

Scale I

Test weight outside tolerance.
Acknowledge with
[START/RESET] key.

Scale I

Zero offset outside tolerance.
Acknowledge with
[START/RESET] key.

The system is not in a stable
position. The scales are subjected
to a draft. Bags with tubing system
are placed on the scales; this is
especially applicable to filtrate
scales.

Lifting magnet for ball not
connected or defective.

Ball mechanics not smooth. Check for smoothness and adjust,

The system is not in a stable
position. The scales are subjected
to a draft. Bags with tubing system
are placed on the scales; this is
especially applicable to filtrate
scales.

Scales mechanics not smooth. Check for smoothness and adjust,

Test weight fails to drop back. Ball
mechanics not smooth.

Set up the system in a stable
position. Remove the system from
the draft, or close the window.
Relieve the scales.

Fit the connector to the
motherboard. If defective, the lifting
magnet must be replaced.

if necessary. Weight check in the
Service program.

Set up the system in a stable
position. Remove the system from
the draft, or close the window.
Relieve the scales.

if necessary. Weight check in the
Service program.

Check for smoothness and adjust,
if necessary. Weight check in the
Service program (ESC Service).

2.4.3 Pump Test

In the first step, the stop-by-OP function is tested, i.e. the pumps must
stop running. In this step, the safety processor releases the pumps, and
the operating processor does not activate the pumps. In the second
step, all of the four pumps are activated with a rate of 100 ml/min. The
safety processor checks whether the pumps are running at the correct
rate and whether the reed contact (line inserting position) is actuated. In
the third step, the operating processor activates the pumps and the
safety processor disables the pump activation, i.e. the pumps may not
be running. In the event of an error, messages are emitted. Here, the
error messages section only relates to the blood pump. The messages
are identical for the other pumps, i.e. the filtrate, substituate and
dialysate pumps.

2-6 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

Error message Causes Action required

Blood pump

Blood pump door open.
Acknowledge with
[START/RESET] key.

Blood pump

Not stopped by operating
processor.
Acknowledge with
[START/RESET] key.

Blood pump

Wrong speed, rate.
Acknowledge with
[START/RESET] key.

Blood pump door open. Close the blood pump door.

Magnet in blood pump door
missing.

Hall sensor in pump housing not
connected or defective.

Control electronics (P.C.B. LP 123)
of the specified pump defective.

P.C.B. LP 244-SR defective. Replace P.C.B. LP 244-SR.

Control electronics (P.C.B. LP 123)
of the specified pump defective.

P.C.B. LP 244-SR defective. Replace P.C.B. LP 244-SR.

Clock pulse generator of motor on
P.C.B. LP 123 not connected.

Control electronics (P.C.B. LP 123)
of the specified pump defective.

Replace the blood pump door.

Plug the connector onto the
appropriate P.C.B. LP 123 or
replace the pump housing.

Replace P.C.B. LP 123.

Caution: The BLD must be
calibrated.

Replace P.C.B. LP 123.

Caution: The BLD must be
calibrated.

Connect the clock pulse generator
socket.

Replace P.C.B. LP 123.

Blood pump

Reed contact  line threading
position
Acknowledge with
[START/RESET] key.

Defect of the clock pulse generator
on the geared motor.

Magnet in rotor missing, defective. Replace the rotor.

Reed switch in pump housing not
connected or defective.

Control electronics (P.C.B. LP 123)
of the specified pump defective.

P.C.B. LP 244-SR defective. Replace P.C.B. LP 244-SR.

Replace the geared motor.

Caution: Blood pump
20 V/3300 rpm; all other pumps
24 V/3300 rpm.

Plug the connector onto the
appropriate P.C.B. LP 123 or
replace the pump housing.

Replace P.C.B. LP 123.

Caution: The BLD must be
calibrated.

Fresenius Medical Care multiFiltrate TM 6/03.07 2-7

Chapter 2: Functional Description

Error message Causes Action required

Blood pump

Stop.
Acknowledge with
[START/RESET] key.

Blood pump

Not stopped by safety processor.
Acknowledge with
[START/RESET] key.

2.4.4 Pressure Transducer

+24-V supply voltage missing. Check the fuse on the PSU board

LP 128. If necessary, replace
P.C.B. LP 128.

Caution: All pressures must be
checked and calibrated, if
necessary.

Control electronics (P.C.B. LP 123)
of the specified pump defective.

P.C.B. LP 244-OP defective.
(The pumps are activated serially,
i.e. the operating processor is
defective only if all pumps fail to be
activated.)

Control electronics (P.C.B. LP 123)
of the specified pump defective.

P.C.B. LP 244-SR defective. Replace P.C.B. LP 244-SR.

Replace P.C.B. LP 123.

Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

Replace P.C.B. LP 123.

Caution: The BLD must be
calibrated.

In the first step, the zero point, amplification and detuning factors are
tested. In the second step, the zero points of the pressure transducers
are checked with a tolerance of ±20 mmHg. In the third step, Part and
Pven are detuned to 300 mmHg. The values are checked with a
tolerance of ±20 mmHg. In the fourth step, PPHF and PFil are detuned
to 300 mmHg. The values are checked with a tolerance of ±20 mmHg.
In the event of an error, messages are emitted. The following table only
shows the error messages for «arterial pressure». The messages are
identical for the venous, pre-hemofilter and filtrate pressures.

Error message Causes Action required

Arterial pressure

Factors / Offset lost.
Acknowledge with
[START/RESET] key.

Lithium battery on P.C.B. LP 244­OP discharged.

Replace the battery and execute
the calibration step in the Service
program.

2-8 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

Error message Causes Action required

Arterial pressure

Zero outside tolerance.
Acknowledge with
[START/RESET] key.

Arterial pressure

Detuning outside tolerance
Acknowledge with
[START/RESET] key.

Zero point drifted off. Check the setting and calibrate the

pressure in the Service program.

P.C.B. LP 343-1 (for venous
pressure: P.C.B. LP 450-3)
defective.

P.C.B. LP 244-OP defective. Replace P.C.B. LP 244-OP.

Amplification drifted off. Check the setting and calibrate the

P.C.B. LP 343-1 (for venous
pressure: P.C.B. LP 450-3)
defective.

Replace P.C.B. LP 343-1 (LP 450-

3) and calibrate the pressure in the
Service program.

Caution: When replacing the
P.C.B. LP 450-3, check / adjust the
air detector, optical detector and
venous pressure transducer.

Caution: All scales and pressures
must be calibrated.

pressure in the Service program.

Replace P.C.B. LP 343-1 (LP 450-

3) and calibrate the pressure in the
Service program.

Caution: When replacing the
P.C.B. LP 450-3, check / adjust the
air detector, optical detector and
venous pressure transducer.

2.4.5 Optical Detector

P.C.B. LP 244-OP defective. Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

In the first step, the non-opaque state is checked. In the second step,
the optical detector is detuned and the opaque state is checked. In the
event of an error, messages are emitted.

Fresenius Medical Care multiFiltrate TM 6/03.07 2-9

Chapter 2: Functional Description

Error message Causes Action required

Opt. detector

Senses opaque fluid.
Acknowledge with
[START/RESET] key.

Opt. detector

Fails to sense opaque fluid after
attenuation.
Acknowledge with
[START/RESET] key.

Blood present in the system or
objects inserted in the OD.

OD adjusted improperly. Check the setting of the OD.

P.C.B. LP 450-3 defective. Replace the P.C.B. LP 450-3.

Measuring head of OD defective. Replace and adjust the OD

P.C.B. LP 244-SR defective. Replace P.C.B. LP 244-SR.

OD adjusted improperly. Check the setting of the OD.

P.C.B. LP 450-3 defective. Replace the P.C.B. LP 450-3.

P.C.B. LP 244-OP defective. Replace P.C.B. LP 244-OP.

Remove the blood line or the
objects from the OD.

Caution: When replacing the
P.C.B. LP 450-3, check / adjust the
air detector, optical detector and
venous pressure transducer.

measuring head.

Caution: The BLD must be
calibrated.

Caution: When replacing the
P.C.B. LP 450-3, check / adjust the
air detector, optical detector and
venous pressure transducer.

2.4.6 Air Detector

Caution: All scales and pressures

must be calibrated.

AD 28 defective. Replace AD 28-1.

In the first step, it is checked whether fluid is present in the drip
chamber. if YES, the clamp is activated (opened). In the second step,
the air detector is detuned and the alarm state checked. In the event of
an error, messages are emitted.

2-10 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

Error message Causes Action required

Air detector

LDA1  not in alarm mode.
Acknowledge with
[START/RESET] key.

Air detector

Clamp does not close.
Acknowledge with
[START/RESET] key.

LD adjusted improperly. Check the setting of the LD.

P.C.B. LP 450-3 defective. Replace the P.C.B. LP 450-3.

Caution: When replacing the
P.C.B. LP 450-3, check / adjust the
air detector, optical detector and
venous pressure transducer.

Ultrasonic detector defective. Replace and adjust the ultrasonic

detector.

P.C.B. LP 244-SR defective. Replace P.C.B. LP 244-SR.

Caution: The BLD must be
calibrated.

LD adjusted improperly (LDA2). Check the setting of the LD.

P.C.B. LP 450-3 defective. Replace the P.C.B. LP 450-3.

Caution: When replacing the
P.C.B. LP 450-3, check / adjust the
air detector, optical detector and
venous pressure transducer.

P.C.B. LP 244-SR defective. Replace P.C.B. LP 244-SR.

Caution: The BLD must be
calibrated.

2.4.7 Blood Leak Detector

In the first step, the blood leak detector is checked for being in an
acceptable state. In the second step, the blood leak detector is detuned
and the alarm state checked. In the event of an error, messages are
emitted.

Error message Causes Action required

Blood leak detector

Calibration values incorrect
or missing.
Acknowledge with
[START/RESET] key.

Lithium battery on P.C.B. LP 244­SP discharged.

Replace the battery and execute
the calibration step in the Service
program.

Fresenius Medical Care multiFiltrate TM 6/03.07 2-11

Chapter 2: Functional Description

Error message Causes Action required

Blood leak detector

Outside acceptable range.
Remove filtrate line.
Acknowledge with
[START/RESET] key.

Blood leak detector

Alarm-free after signal attenuation.
Acknowledge with
[START/RESET] key.

2.4.8 Heater

Empty filtrate line or objects
inserted in the BLD sensor head.

P.C.B. LP 125 and/or BLD sensor
head defective.

P.C.B. LP 244-SR defective. Replace P.C.B. LP 244-SR.

P.C.B. LP 125 and/or BLD sensor
head defective.

P.C.B. LP 244-OP defective. Replace P.C.B. LP 244-OP.

In the first step, the safety relay is checked for being in the open state.
In addition, the sensors are checked for interruption and short-circuit. In
the second step, the safety relay is activated and checked for being in
the closed state. In the third step, the bag and foil sensors of OP and
SP are checked for being synchronized. Tolerances are ±1 °C. In the
fourth step, the foil sensor is detuned to >120 °C and, thus, the safety
shutoff mechanism (fire protection) is checked. In the fifth step, the bag
and foil sensors are detuned to >41 °C to check whether the heater is
switched off in case of an overtemperature of the solutions. In the sixth
step, it is checked whether the heater can be activated. In the event of
an error, messages are emitted. Here, the error messages section only
relates to the lower heater. The messages are identical for the upper
heater.

Remove the filtrate line or the
objects.

Replace P.C.B. LP 125 and the
BLD sensor head, and calibrate the
BLD.

Caution: The BLD must be
calibrated.

Replace P.C.B. LP 125 and the
BLD sensor head, and calibrate the
BLD.

Caution: All scales and pressures
must be calibrated.

Error message Causes Action required

Lower heater

Switch-off path SP defective
(Sub. relay not open)
Voltage missing.
Fuse defective.
Acknowledge with
[START/RESET] key.

2-12 Fresenius Medical Care multiFiltrate TM 6/03.07

Safety relay on P.C.B. LP 122
defective.

Activation on P.C.B. LP 244-SR
defective.

Supply voltage for heater not
connected.

Fuse on P.C.B. LP 122 defective. Replace the fuse.

Replace P.C.B. LP 122.

Replace P.C.B. LP 244-SR.

Caution: The BLD must be
calibrated.

Fit the socket on P.C.B. LP 122.

Chapter 2: Functional Description

Error message Causes Action required

Lower heater

Control FET short-circuit
Acknowledge with
[START/RESET] key.

Lower heater

Bag sensor broken.
Acknowledge with
[START/RESET] key.

Lower heater

Foil sensor broken.
Acknowledge with
[START/RESET] key.

Lower heater

Control FET defective. Replace P.C.B. LP 122.

Activation on P.C.B. LP 244-OP
defective.

Socket of sensors not fitted. Fit the socket on P.C.B. LP 122.

Sensor interrupted. Replace the heater.

P.C.B. LP 122 defective Replace P.C.B. LP 122.

Converter on P.C.B. LP 244-SR
defective.

Socket of sensors not fitted. Fit the socket on P.C.B. LP 122.

Sensor interrupted. Replace the heater.

P.C.B. LP 122 defective Replace P.C.B. LP 122.

Converter on P.C.B. LP 244-SR
defective.

Bag sensor short-circuit. Replace the heater.

Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

Replace P.C.B. LP 244-SR.

Caution: The BLD must be
calibrated.

Replace P.C.B. LP 244-SR.

Caution: The BLD must be
calibrated.

Bag sensor short-circuit.
Acknowledge with
[START/RESET] key.

Lower heater

Foil sensor short-circuit.
Acknowledge with
[START/RESET] key.

Lower heater

Sub. relay not closed,
foil defective.
Acknowledge with
[START/RESET] key.

P.C.B. LP 122 defective Replace P.C.B. LP 122.

Converter on P.C.B. LP 244-SR
defective.

Bag sensor short-circuit. Replace the heater.

P.C.B. LP 122 defective Replace P.C.B. LP 122.

Converter on P.C.B. LP 244-SR
defective.

Supply voltage for heater not
connected.

Fuse on P.C.B. LP 122 defective. Replace the fuse.

Heater foil not connected. Connect the heater foil to P.C.B.

Heater foil interrupted. Replace the heater.

Safety relay defective. Replace P.C.B. LP 122.

Activation on P.C.B. LP 244-SR
defective.

Replace P.C.B. LP 244-SR.

Caution: The BLD must be
calibrated.

Replace P.C.B. LP 244-SR.

Caution: The BLD must be
calibrated.

Fit the socket on P.C.B. LP 122.

LP 122.

Replace P.C.B. LP 244-SR.

Caution: The BLD must be
calibrated.

Fresenius Medical Care multiFiltrate TM 6/03.07 2-13

Chapter 2: Functional Description

Error message Causes Action required

Lower heater

Bag sensors,
OP-SP not synchronous
Acknowledge with
[START/RESET] key.

Lower heater

Foil sensors,
OP-SP not synchronous
Acknowledge with
[START/RESET] key.

Heater still warm after last
treatment or Service program.

Sensor board (P.C.B. LP 1220)
defective.

P.C.B. LP 122 defective Replace P.C.B. LP 122.

Converter on P.C.B. LP 244-SR
defective.

Converter on P.C.B. LP 244-OP
defective.

Heater still warm after last
treatment or Service program.

Sensor board (P.C.B. LP 1220)
defective.

P.C.B. LP 122 defective Replace P.C.B. LP 122.

Converter on P.C.B. LP 244-SR
defective.

Allow the heater to cool down.

Replace the heater.

Replace P.C.B. LP 244-SR.

Caution: The BLD must be
calibrated.

Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

Allow the heater to cool down.

Replace the heater.

Replace P.C.B. LP 244-SR.

Caution: The BLD must be
calibrated.

Lower heater

Foil sensor detuning,
no overtemperature cutoff.
Acknowledge with
[START/RESET] key.

Lower heater

Bag sensor detuning,
no overtemperature.
Acknowledge with
[START/RESET] key.

Converter on P.C.B. LP 244-OP
defective.

DIL relay on P.C.B. LP 122
defective.

DIL relay on P.C.B. LP 122
defective.

Activation on P.C.B. LP 244-OP
defective.

Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

Replace P.C.B. LP 122.

Replace P.C.B. LP 122.

Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

2-14 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

Error message Causes Action required

Lower heater

Foil sensor detuning,
no overtemperature.
Acknowledge with
[START/RESET] key.

Lower heater

FET control defective
Acknowledge with
[START/RESET] key.

DIL relay on P.C.B. LP 122
defective.

Activation on P.C.B. LP 244-OP
defective.

FET on P.C.B. LP 122 defective. Replace P.C.B. LP 122.

Activation on P.C.B. LP 44-OP
defective.

Replace P.C.B. LP 122.

Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

2.4.9 Battery Test, Part 2

The first step was carried out at the beginning of the functional test, with
acceptance of the starting value and loading of the battery. In the
second step, the voltage is checked after loading and is compared with
the starting value. In the event of an error, messages are emitted.

Error message Causes Action required

Battery test

Battery not connected, defective.
Acknowledge with
[START/RESET] key.

Battery not connected. Connect the rechargeable battery.

Battery terminal voltage without
load <<18 V.

Converter on P.C.B. LP 244-OP
defective.

Replace the battery.

Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

Battery test

Load test failed.
Acknowledge with
[START/RESET] key.

Fresenius Medical Care multiFiltrate TM 6/03.07 2-15

Resistor or load relay on the PSU,
P.C.B. LP 128, defective.

Converter or activation on P.C.B.
LP 244-OP defective.

Replace P.C.B. LP 128.

Caution: All pressures must be
checked and calibrated, if
necessary.

Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

Chapter 2: Functional Description

Error message Causes Action required

Battery test

Insufficient capacity.
Acknowledge with
[START/RESET] key.

System not used for a prolonged
period or frequent power failure
(preparation).

Battery defective.

Charging circuit on the PSU board,
P.C.B. LP 128, defective.

Converter on P.C.B. LP 244-OP
defective.

Load the battery from the mains for
a minimum of 10 hours.

Replace the battery.

Replace P.C.B. LP 128.

Caution: All pressures must be
checked and calibrated, if
necessary.

Replace P.C.B. LP 244-OP.

Caution: All scales and pressures
must be calibrated.

2.4.10 Audible Alarm

In the first step, it is checked whether the audible alarm is silenced. In
the second step, the audible alarm is generated and checked for proper
functioning by means of a microphone. In the event of an error,
messages are emitted.

Error message Causes Action required

Audible alarm

Not silenced.
Acknowledge with
[START/RESET] key.

Constant ambient noise. Repeat the test after the noise has

stopped.

Loud humming noise on the
loudspeaker.

Check the route of the loudspeaker
cable. Replace the loudspeaker­microphone unit or the
motherboard P.C.B. LP 124.

The microphone line is disturbed. Check the route of the microphone

cable.

Microphone or amplifier defective. Replace the loudspeaker-

microphone unit or the
motherboard P.C.B. LP 124.

Audible alarm

Not active.
Acknowledge with
[START/RESET] key.

2-16 Fresenius Medical Care multiFiltrate TM 6/03.07

Alarm tone not audible. Loudspeaker cable connected to

P.C.B. LP 124?

Replace the loudspeaker­microphone unit or the
motherboard P.C.B. LP 124.

Error message despite audible
alarm.

Microphone cable connected to
P.C.B. LP 124?

Replace the loudspeaker­microphone unit or the
motherboard P.C.B. LP 124.

Chapter 2: Functional Description

2.4.11 Heparin Pump

After turning power on to the multiFiltrate, a self-test of the heparin
pump will be performed. Immediately before the multiFiltrate self-test,
the operating processor will query the heparin pump status. In the event
of an error, messages are emitted. If the heparin pump is not installed,
not connected to the power supply or deactivated via Service option, no
error message will be displayed.

Error message Causes Action required

Heparin pump

Heparin pump not ready.
No communication.

Heparin pump

Heparin pump optionally
deactivated in Service mode

Heparin pump

Heparin pump detects internal error

Heparin pump

Wrong / unauthorized syringe type
detected

Heparin pump

Unknown hardware error

Heparin pump does not respond. Acknowledge with

[START/RESET] key.

Heparin pump deactivated.

Heparin pump manually
deactivated in the Service mode

Internal heparin pump error without
definite cause.

Wrong syringe type set. Acknowledge with

Heparin pump detects an unknown
hardware error.

Acknowledge with
[START/RESET] key.

Heparin pump deactivated.

Acknowledge with
[START/RESET] key.

Heparin pump deactivated.

[START/RESET] key.

Heparin pump deactivated.

Select a valid syringe type.
0 = 50 ml P syringe

1 = 30 ml heparin syringe
2 = 50 ml Injectomat syringe

Acknowledge with
[START/RESET] key.

Heparin pump deactivated.

Heparin pump

Heparin pump hardware error
Gate array error

Heparin pump

Spike on reset line

Heparin pump

Wrong HEX switch position

Heparin pump

Powerdown

Fresenius Medical Care multiFiltrate TM 6/03.07 2-17

Heparin pump detects an error in
the gate array.

A reset occurred during operation. Acknowledge with

Wrong HEX switch position set. Acknowledge with

Powerdown without 24 V-cutoff. Acknowledge with

Acknowledge with
[START/RESET] key.

Heparin pump deactivated.

[START/RESET] key.

Heparin pump deactivated.

[START/RESET] key.

Heparin pump deactivated.

Set a valid HEX switch position.

[START/RESET] key.

Heparin pump deactivated.

Chapter 2: Functional Description

Error message Causes Action required

Heparin pump

Operating processor internal
communication error with heparin
pump

Heparin pump

Watchdog error

Heparin pump

NOVRAM error

Heparin pump

CAMUS transmission error

2.4.12 multiDataLink

Erroneous data transmission.
Wrong or missing characters.
Check sum error.

The watchdog is not able to
interrupt the 24 V control voltage
for the stepper motor.

The CRC protection of the data
saved in the E2PROM is not
correct.

Not used in the multiFiltrate. Not used in the multiFiltrate.

Immediately before the multiFiltrate self-test, the operating processor
will query the multiDataLink version identifications. A response given by
the mDL is an indicator for its operability. In the event of an error,
messages are emitted. If the multiDataLink is not installed, not
connected to the power supply or deactivated via Service option, no
error message will be displayed.

Acknowledge with
[START/RESET] key.

Heparin pump deactivated.

Acknowledge with
[START/RESET] key.

Heparin pump deactivated.

Acknowledge with
[START/RESET] key.

Heparin pump deactivated.

Error message Causes Action required

multiDataLink

multiDataLink not standby
No communication

multiDataLink

multiDataLink optionally
deactivated in Service mode

multiDataLink does not respond. Acknowledge with

[START/RESET] key.

multiDataLink deactivated.

multiDataLink manually
deactivated in the Service program.

Acknowledge with
[START/RESET] key.

multiDataLink deactivated.

Activate multiDataLink in the
Service program.

2.4.13 Ci-Ca Module (Option)

The test for the Ci-Ca module can not be enabled if the processor test
of the module and the multiFiltrate was not passed successfully. If these
processor tests are performed successfully, the T1 test will be started
automatically as soon as the operator has confirmed that the starting
conditions are met.

The following functions of the Ci-Ca module are tested:

2-18 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

 Stop of the citrate and the calcium pump by the module’s operating

processor
 Stop of the citrate and the calcium pump by the system’s watchdog.
 Function of the citrate pump and the insertion switch
 Function of the calcium pump and the insertion switch

If a test could not be passed successfully, a warning indicating an error
number appears on the multiFiltrate display. The test can be repeated
any number of times, however, it cannot be skipped.

The T1 test of the multiFiltrate system is performed simultaneously to
the T1 test of the Ci-Ca module.

Error
number

281 Lacking Ci Hall impulse after test was started

282 Ci Hall impulse too early during pump stop test

283 Ci Hall impulse too late during pump stop test

285 Ci Hall impulse too early during pump stop test

286 Ci Hall impulse too late during pump stop test

288 Ci Hall impulse too early during insertion switch

289 Ci Hall impulse too late during insertion switch

291 Lacking Ca Hall impulse after test was started

292 Ca Hall impulse too early during pump stop test

Description Possible cause

(>5 s).

(<7 s).

(>9 s).

(<7 s).

(>9 s).

test (<7 s).

test (>9 s).

(>5 s).

(<7 s).

The rotor of the citrate pump is loose, jammed or
blocked.

Insertion switch pressed —> citrate line already
inserted.

The rotor of the citrate pump is loose, jammed or
blocked.

Insertion switch pressed —> citrate line already
inserted during the test.

The rotor of the citrate pump is loose, jammed or
blocked.

Insertion switch pressed —> citrate line already
inserted during the test.

The rotor of the citrate pump is loose, jammed or
blocked.

The rotor of the Ca pump is loose, jammed or
blocked.

Insertion switch pressed —> calcium line already
inserted.

293 Ca Hall impulse too late during pump stop test

(>9 s).

295 Ca Hall impulse too early during pump stop test

(<7 s).

296 Ca Hall impulse too late during pump stop test

(>9 s).

298 Ca Hall impulse too early during insertion switch

test (<7 s).

299 Ca Hall impulse too late during insertion switch

test (>9 s).

Fresenius Medical Care multiFiltrate TM 6/03.07 2-19

The rotor of the Ca pump is loose, jammed or
blocked.

Insertion switch pressed —> calcium line already
inserted during the test.

The rotor of the Ca pump is loose, jammed or
blocked.

Insertion switch pressed —> calcium line already
inserted during the test.

The rotor of the Ca pump is loose, jammed or
blocked.

Chapter 2: Functional Description

2.5 Error Messages

2.5.1 Alarm Messages

These messages cause all pumps to stop

Code Description

E10 Blood pump: Blood pump door open

E11 Blood pump: Stop

E12 Blood pump: Wrong speed, rate

E13 Arterial pressure, arterial pressure too low

E14 Arterial pressure, arterial pressure too high

E15 Venous pressure, venous pressure too low

E16 Venous pressure, venous pressure too high

E17 TMP, TMP too low

E18 TMP, TMP too high

E19 System stopped, Stop key has been pressed

E20 Pre-filter pressure/ Pre-filter pressure too low (for

hemoperfusion)

E21 Pre-filter pressure, pre-filter pressure too low

E22 Pre-filter pressure, pre-filter pressure too high

E23 Non-opaque/opaque fluid detector, venous detector

senses non-opaque or opaque fluid

E24 Air detector, air and/or microbubbles

E25 Blood leak detector, blood leak, hemolysis, membrane

rupture

E26 Blood leak detector, 2 min override active

E27 Power failure, emergency operation available for max.

15 min.

Converted by OP from SP code 130.

E28 End of emergency operation, the unit will turn itself off

automatically after 5 minutes.

Converted by OP from SP code 189.

E29 Pre-filter pressure/ Pre-filter pressure too high (for

hemoperfusion)

E30 Blood lines: Blood line jammed / Hydrophobic filter

P_pHF wet

E31 Blood lines: Rotor broken

2-20 Fresenius Medical Care multiFiltrate TM 6/03.07

Code Description

Chapter 2: Functional Description

E32 Blood lines: Hydrophobic filter P

E33 Blood lines: Hydrophobic filter P

art

ven

wet

wet

E34 Air detector: Transmitting voltage on the air detector too

high

E35 Ci-Ca module: Citrate pump turning too slowly (volume

monitoring)

E36 Ci-Ca module: Citrate pump turning too fast (volume

monitoring)

E37 Ci-Ca module: Calcium pump turning too slowly

(volume monitoring)

E38 Ci-Ca module: Calcium pump turning too fast (volume

monitoring)

E39 Ci-Ca module: Dialysis is started although the primed

Ci/Ca lines are not free from air

E40 Ci-Ca module: Drop counter citrate  insufficient

number of drops; check citrate bag, drip chamber and
tubing system!

E41 Ci-Ca module: Number of drops counted by citrate drop

counter too high; check citrate drip chamber and tubing
system!

E42 Ci-Ca module: Drop counter calcium  insufficient

number of drops; check calcium bag, drip chamber and
tubing system!

E43 Ci-Ca module: Number of drops counted by calcium

drop counter too high; check calcium drip chamber and
tubing system!

E44 Ci-Ca module: Citrate pump turning too slowly; (motor

steps)

E45 Ci-Ca module: Citrate pump turning too fast; (motor

steps)

E46 Ci-Ca module: Calcium pump turning too slowly (motor

steps)

E47 Ci-Ca module: Calcium pump turning too fast (motor

steps)

E48 Ci-Ca module: Citrate insertion switch without

contact,Ci pump segment inserted properly?

E49 Ci-Ca module: Calcium insertion switch without

contact,Ca pump segment inserted properly?

E50 Ci-Ca module: The citrate bag change takes more than

2 min; finish bag change!

Fresenius Medical Care multiFiltrate TM 6/03.07 2-21

Chapter 2: Functional Description

Code Description

E51 Ci-Ca module: The Ca bag change takes more than

2 min; finish bag change!

E52-54 Not used

E55 Ci-Ca module: MPS treatment is permitted for a

maximum of 4 hours!; initiate reinfusion!

E56 Ci-Ca module: HP treatment is permitted for a

maximum of 4 hours!; initiate reinfusion!

E57 Ci-Ca module: Balance switched off; Ca supply

interrupted!

E58 Ci-Ca module: System error of Ci-Ca module:

Anitcoagulation with citrate completed!

E59 Ci-Ca module: Drop counter citrate  insufficient

number of drops; Ci bag change required?

E60 Ci-Ca module: Drop counter calcium  insufficient

number of drops; Ca bag change required?

2.5.2 Warning Messages

These messages cause the balancing to stop. The blood circuit will be
maintained.

For error messages no. 132 through 144 the scales must be
recalibrated.

Code Description

W62 Blood pump: Blood pump door open

W63 Filtrate pump: Cover open

W64 Sub pump: Cover open

W65 Dialysate pump: Cover open

W66 Prime mode: No weight loss on sub-scale-post

(HVCVVH)

W67 Prime mode: No weight loss on sub-scale-pre

(HVCVVH)

W68 Bag change: Overload on sub-scale-post (HVCVVH)

W69 Bag change: Sub-scale post underload (HVCVVH)

W70 Opt. detector: is opaque

W71 Opt. detector: Fails to sense opaque fluid after

attenuation

W72 Scale 1: Temp. coeff. factors missing, incorrect

2-22 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

Code Description

W73 Scale 2: Temp. coeff. factors missing, incorrect

W74 Scale 3: Temp. coeff. factors missing, incorrect

W75 Scale 4: Temp. coeff. factors missing, incorrect

W76 Scale 1: Calibration factor missing, incorrect

W77 Scale 2: Calibration factor missing, incorrect

W78 Scale 3: Calibration factor missing, incorrect

W79 Scale 4: Calibration factor missing, incorrect

W80 Scale 1: Movement detected on Scale 1

W81 Scale 2: Movement detected on Scale 2

W82 Scale 3: Movement detected on Scale 3

W83 Scale 4: Movement detected on Scale 4

W84 Scale 1: Test weight outside tolerance

W85 Scale 2: Test weight outside tolerance

W86 Scale 3: Test weight outside tolerance

W87 Scale 4: Test weight outside tolerance

W88 Scale 1: Test weight fails to drop back

W89 Scale 2: Test weight fails to drop back

W90 Scale 3: Test weight fails to drop back

W91 Scale 4: Test weight fails to drop back

W92 Air detector: LDA1  not in alarm mode

W93 Air detector: Clamp fails to close

W94 Lower heater: Switch-off path SP defective (Sub-relay

not open)

W95 Upper heater: Switch-off path SP defective (dial. relay

not open)

W96 Lower heater: Control FET short-circuit

W97 Upper heater: Control FET short-circuit

W98 Lower heater: Bag sensor broken

W99 Lower heater: Foil sensor broken

W100 Upper heater: Bag sensor broken

W101 Upper heater: Foil sensor broken

W102 Lower heater: Bag sensor short-circuit

W103 Lower heater: Foil sensor short-circuit

Fresenius Medical Care multiFiltrate TM 6/03.07 2-23

Chapter 2: Functional Description

Code Description

W104 Upper heater: Bag sensor short-circuit

W105 Upper heater: Foil sensor short-circuit

W106 Lower heater: Sub. relay not closed, foil defective

W107 Upper heater: Dial. relay not closed, foil defective

W108 Lower heater: Foil sensor detuning, no over-

temperature cutoff

W109 Upper heater: Foil sensor detuning, no over-

temperature cutoff

W110 Lower heater: Bag sensors, OP-SP not synchronous

W111 Upper heater: Bag sensors, OP-SP not synchronous

W112 Lower heater: Bag sensor detuning, no

overtemperature

W113 Lower heater: Foil sensor detuning, no overtemperature

W114 Upper heater: Bag sensor detuning, no

overtemperature

W115 Upper heater: Foil sensor detuning, no overtemperature

W116 Lower heater: FET control defective

W117 Upper heater: FET control defective

W118 Battery test: Battery not connected, defective

W119 Battery test: Load test failed

W120 Battery test: Insufficient capacity

W121 Blood leak detector: Outside acceptable range, remove

filtrate line

W122 Blood leak detector: Alarm-free after attenuation

W123 Prime mode: No weight loss on SUB scale

W124 Prime mode: No weight loss on DIA scale

W125 Prime mode: No weight loss on SUB scale (CVVHDF)

W126 Prime mode: No weight loss on DIA scale (CVVHDF)

W127 Blood leak detector: Calibration values missing,

incorrect

W128 Audible alarm: Not silenced

W129 Audible alarm: not active

W130 Power failure: Emergency operation available for max.

15 minutes (OP generates code 27)

2-24 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

Code Description

W131 Incorrect allocation of scales, repeat Preparation (very

unusual)

W132 Load on scale, incorrect load on scale I (initial weight)

W133 Load on scale, incorrect load on scale II (initial weight)

W134 Load on scale, incorrect load on scale III (initial weight)

W135 Load on scale, incorrect load on scale IV (initial weight)

W136 Balancing, Scale I, leakage from tubing system, objects

W137 Balancing, Scale II, leakage from tubing system, objects

W138 Balancing, upper scales, leakage from tubing system,

objects

W139 Balancing, filtrate bag full, replace or empty filtrate bag

W140 Balancing, filtrate scales moved, leakage from tubing

system

W141 Scale 1, overload or underload, check load, scale

jammed?

W142 Scale 2, overload or underload, check load, scale

jammed?

W143 Scale 3, overload or underload, check load, scale

jammed?

W144 Scale 4, overload or underload, check load, scale

jammed?

W145 Bag change: Overload on SUB scale (CVVHDF)

W146 Bag change: Overload on SUB scale 1 (CVVH, HF)

W147 Bag change: Overload on SUB scale 2 (CVVH, HF)

W148 Bag change: SUB on wrong scale (CVVHDF)

W149 Bag change: Insufficient substituate (CVVH, HF)

W150 Bag change: Insufficient substituate (CVVHDF)

W151 Bag change: Insufficient saline solution on scale 1

(MPS)

W152 Bag change: Overload on plasma scale 1 (MPS)

W153 Bag change: Overload on plasma scale 2 (MPS)

W154 Bag change: Overload on dialysate scale (CVVHDF)

W155 Bag change: Overload on dialysate scale 1 (CVVHD)

W156 Bag change: Overload on dialysate scale 2 (CVVHD)

W157 Dialysate on wrong scale (tubing arrangement test after

bag change HDF)

Fresenius Medical Care multiFiltrate TM 6/03.07 2-25

Chapter 2: Functional Description

Code Description

W158 Bag change: Insufficient dialysate (CVVHD)

W159 Bag change: Insufficient dialysate (CVVHDF)

W160 Bag change: Overload on filtrate scale

W161 TMP/MPS TMP >100mmHg

W162 Lower heater: MPS temperature > 37°C

W163 Upper heater: MPS temperature > 37°C

W164 Lower heater: Temperature too high

W165 Upper heater: Temperature too high

W166 Bag change: Overload SUB pre (HVCVVH)

W167 Bag change: Insufficient SUB pre (HVCVVH)

W168 Balancing: Ultrafiltration or substitution rate too high

W169 Lower heater: Foil sensors, OP-SP not synchronous

W170 Upper heater: Foil sensors, OP-SP not synchronous

W171 Balancing: Aborted because maximum allowable

deviation of 500g was exceeded

W172 Lower heater: MPS foil temperature too high, risk of

hemolysis

W173 Upper heater: MPS foil temperature too high, risk of

hemolysis

W174 Balancing: Balancing stopped, balancing was switched

off

W175 Battery capacity: Charging voltage too high, battery

deactivated, no power failure

W176 Filtrate scale: Test weight outside tolerance (T0 test)

W177 Filtrate scale: Test weight fails to drop back (T0 test)

W178 Balancing: UF goal reached, ultrafiltration was set to 0

W179 Balancing: Substituate bag empty, replace substituate

bag

W180 Balancing: Dialysate bag empty, replace dialysate bag

W181 Balancing: Movement detected on filtrate scales,

leakage from tubing system

W182 Filtrate scale: Overload or underload, check load, scale

jammed?

W183 Balancing: Plasma bag empty, open outlet line(s) on

plasma bags

W184 Balancing: Plasma pump delivers from wrong scale (I)

2-26 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

Code Description

W185 Balancing: Plasma pump delivers from wrong scale (II)

W186 Balancing: Sub-postdilution bag empty, replace

substituate bag

W187 Balancing: Sub-predilution bag empty, replace

substituate bag

W188 End of treatment: Plasma infused (MPS only)

W189 End of emergency operation: The unit will turn itself off

automatically after 5 minutes

W190 Power failure end

W191 Bag change: Insufficient plasma on scale 2

W192 Prime mode: Overload on dialysate scale (CVVHDF)

W193 Prime mode: Overload on SUB scale (CVVHDF)

W194 Prime mode: Insufficient substituate (CVVHDF)

W195 Prime mode: Insufficient dialysate (CVVHDF)

W196 Prime mode: Overload on sub-scale-pre (HVCVVH)

W197 Prime mode: Overload on sub-scale-post (HVCVVH)

W198 Prime mode: Insufficient SUB pre (HVCVVH)

W199 Prime mode: Insufficient SUB post (HVCVVH)

W200 Blood pump: Not stopped by operating processor

W201 Filtrate pump: Not stopped by operating processor

W202 Sub pump: Not stopped by operating processor

W203 Dialysate pump: Not stopped by operating processor

W204 Blood pump: Wrong speed, rate

W205 Filtrate pump: Wrong speed, rate

W206 Sub pump: Wrong speed, rate

W207 Dialysate pump: Wrong speed, rate

W208 Blood pump: Reed contact  line threading position

W209 Filtrate pump: Reed contact  line threading position

W210 Sub pump: Reed contact  line threading position

W211 Dialysate pump: Reed contact  line threading position

W212 Blood pump: Stop

W213 Filtrate pump: Stop

W214 Sub pump: Stop

Fresenius Medical Care multiFiltrate TM 6/03.07 2-27

Chapter 2: Functional Description

Code Description

W215 Dialysate pump: Stop

W216 Blood pump: Not stopped by safety processor

W217 Filtrate pump: Not stopped by safety processor

W218 Sub pump: Not stopped by safety processor

W219 Dialysate pump: Not stopped by safety processor

W220 Arterial pressure: Factors / offset missing, incorrect

W221 Venous pressure: Factors / offset missing, incorrect

W222 PHF pressure: Factors / offset missing, incorrect

W223 Filtrate pressure: Factors / offset missing, incorrect

W224 Arterial pressure: Zero outside tolerance

W225 Venous pressure: Zero outside tolerance

W226 PHF pressure: Zero outside tolerance

W227 Filtrate pressure: Zero outside tolerance

W228 Arterial pressure: Detuning outside tolerance

W229 Venous pressure: Detuning outside tolerance

W230 PHF pressure: Detuning outside tolerance

W231 Filtrate pressure: Detuning outside tolerance

W232 Heparin pump: does not respond

W233 Heparin pump: manually deactivated in Service

program

W234 Heparin pump: detects an internal problem

W235 Heparin pump: Wrong/invalid syringe type

W236 Heparin pump: No anticoagulation

W237 Heparin pump: Anticoagulation stopped

W238 Heparin pump: Anticoagulant Used Up

W239 Heparin pump: Syringe not changed

W240 Heparin pump: Non-registered error code

W241 Heparin pump: Gate-array error

W242 Heparin pump: Restart after spike reset

W243 Heparin pump: Hex switch position error

W244 Heparin pump: Powerdown without 24V cutoff

W245 Heparin pump: Slotted disc error

2-28 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

Code Description

W246 Heparin pump: Slotted disc not found

W247 Heparin pump: Communication sumcheck error

W248 Heparin pump: No heparin pump STOP

W249 Heparin pump: Communication error with OP

W250 Heparin pump: Watchdog error

W251 Heparin pump: NOVRAM error

W252 CAMUS transmission error

W253 Not used

W254 Not used

W255 mDL not ready/no communication

W256 mDL deactivated via Service option

W257 The BP rate is < 30 ml/min; no rotor breakage

monitoring; closure of pressure transducer is not
detected.

W259 Lower heater: Switch-off path SP defective (Sub-relay

not open), voltage not present, SI defective (for P.C.B.
LP122-B)

W260 Upper heater: Switch-off path SP defective (dial. relay

not open), voltage not present, SI defective (for P.C.B.
LP122-B)

W261 Lower heater: Control FET short circuit (for P.C.B.

LP122-B)

W262 Upper heater: Control FET short circuit (for P.C.B.

LP122-B)

W263 Air detector: Transmitting voltage for air detector not

present / nonconforming (for AD28)

W264 Ci-Ca module: Ci-Ca module does not respond, select

alternate anticoagulation if necessary!

W265 Ci-Ca module: Switch-off path «sleep» does not react,

select alternate anticoagulation if necessary!

W266 Balancing: HF bag change aborted

W281 Ci-Ca module: Citrate pump — test F1, Rotor slack or

stuck?

W282 Ci-Ca module: Citrate pump — OP test F2, rotor slack?

W283 Ci-Ca module: Citrate pump — OP test F3, rotor slack or

stuck?

W285 Ci-Ca module: Citrate pump — WD test F5, rotor slack?

Fresenius Medical Care multiFiltrate TM 6/03.07 2-29

Chapter 2: Functional Description

Code Description

W286 Ci-Ca module: Citrate pump — WD test F6, rotor slack or

stuck?

W288 Ci-Ca module: Citrate pump — insertion switch test F8,

line pump segment inserted too early or rotor slack?

W289 Ci-Ca module: Citrate pump — insertion switch test F9,

rotor slack or stuck?

W291 Ci-Ca module: Calcium pump test F11, rotor slack or

stuck?

W292 Ci-Ca module: Calcium pump — OP test F12, rotor

slack?

W293 Ci-Ca module: Calcium pump — OP test F13, rotor slack

or stuck?

W295 Ci-Ca module: Calcium pump — WD test F15, rotor

slack?

W296 Ci-Ca module: Calcium pump — WD test F16, rotor slack

or stuck?

2.5.3 Fatal Errors

W298 Ci-Ca module: Calcium pump — insertion switch test

F18, line pump segment inserted too early or rotor
slack?

W299 Ci-Ca module: Calcium pump — insertion switch test

F19, rotor slack or stuck?

W330 Ci-Ca module: Balance switched off!, Ca supply

interrupted!

W331 Ci-Ca module: Filtrate rate too low for set Ca dose!,

Increase dialysate or substituate or Ca dose if
necessary!

W332 Ci-Ca module: Filtrate rate too high for set Ca dose!,

Reduce dialysate or substituate or Ca dose if
necessary!

W333 Ci-Ca module: Blood flow too low for set citrate dose!,

Increase blood flow or citrate dose if necessary!

W334 Ci-Ca module: Blood flow too high for set citrate dose!,

Reduce blood flow or citrate dose if necessary!

These messages cause the treatment to stop.

2-30 Fresenius Medical Care multiFiltrate TM 6/03.07

Chapter 2: Functional Description

Code Description

E500 Scale 1: (factors missing during treatment)

E501 Scale 2: (factors missing during treatment)

E502 Scale 3: (factors missing during treatment)

E503 Scale 4: (factors missing during treatment)

E504 Memory error: (double values incorrect)

E505 Blood pump: (does not respond)

E506 Substituate pump: (does not respond)

E507 Dialysate pump: (does not respond)

E508 Filtrate pump: (does not respond)

E509 Overvoltage +5V

E510 Overvoltage +12V

E511 Processor test OP: CPU defective,

E512 Processor test OP: Incorrect CRC interrupt vectors

E513 Processor test OP: CRC CRC area defective

E514 Processor test OP: CRC Code area defective

E515 Processor test OP: RAM, bank B defective

E516 Processor test OP: RAM, bank 8 defective

E517 State transition

E518 Program sequence

E519 Self-Test

E520 Balancing

E521 SP does not respond

E522 DP does not respond

E523 Program flow OP: Watchdog

E524 Processor test OP: Watchdog does not respond

E525 Processor test SP: CPU

E526 Processor test SP: Program memory  vectors

E527 Processor test SP: Program memory — CRC area

E528 Processor test SP: Program memory — code area

E529 Processor test SP: RAM — bank B

E530 Processor test SP: RAM — bank B

E531 Processor test SP: Watchdog does not respond

Fresenius Medical Care multiFiltrate TM 6/03.07 2-31

Chapter 2: Functional Description

Code Description

E532 Invalid software combination: OP, DP, SP

E533 Ci-Ca module: Logoff of the option by the MFT

E534 Ci-Ca module: Alarm of the option not followed by MFT

alarm

E535 Ci-Ca module: Timeout of the status message of the

MFT (>2.5 s)

E536 Ci-Ca module: Too many status messages of the MFT

(2x<0.5 s)

E537 Ci-Ca module: Receipt of message not confirmed

E538 Ci-Ca module: unknown identifier

E539 Ci-Ca module: Write buffer full

E540 Ci-Ca module: Read buffer full

E541 Ci-Ca module: Unable to write on TXe

E542 Ci-Ca module: Missing setup values from the MFT (at

end of line installation)

E550 Ci-Ca module: Too many extra Hall pulses from the Ca

pump

E551 Ci-Ca module: Ca pump could not be stopped

E560 Ci-Ca module: Too many extra Hall pulses from the

citrate pump

E561 Ci-Ca module: Citrate pump could not be stopped

E570 Ci-Ca module: Error on checking variable replication

(Int)

E571 Ci-Ca module: Error on checking variable replication

(Float)

E572 Ci-Ca module: Error on checking variable replication

(Long)

E573 Ci-Ca module: Error on checking software diversity

2-32 Fresenius Medical Care multiFiltrate TM 6/03.07

2.5.4 Error Codes of Scales Lo-Level Routines

Scale no. 1

Code Description

E30001 Standstill not possible (timeout after 10 sec)

E30002 Weight value out of range (overload or underload)

E30003 Incorrect calibration weight value

E30004 Negative initial weight

E30005 Wrong initial weight

E30006 ACT_WGT 65535

E30007 Alarm limits permanently violated

E30008 INF_WGT invalid

E30050 Calibration factor missing or incorrect

Chapter 2: Functional Description

Scale no. 2

E30051 Temp. coeff. factors missing or incorrect

E30052 Double fuse defective (zero value)

E30053 Temperature sensor missing / defective

Code Description

E30101 Standstill not possible (timeout after 10 sec)

E30102 Weight value out of range (overload or underload)

E30103 Incorrect calibration weight value

E30104 Negative initial weight

E30105 Wrong initial weight

E30106 ACT_WGT 65535

E30107 Alarm limits permanently violated

E30108 INF_WGT invalid

E30150 Calibration factor missing or incorrect

E30151 Temp. coeff. factors missing or incorrect

E30152 Double fuse defective (zero value)

E30153 Temperature sensor missing / defective

Fresenius Medical Care multiFiltrate TM 6/03.07 2-33

Chapter 2: Functional Description

Scale no. 3

Code Description

E30201 Standstill not possible (timeout after 10 sec)

E30202 Weight value out of range (overload or underload)

E30203 Incorrect calibration weight value

E30204 Negative initial weight

E30205 Wrong initial weight

E30206 ACT_WGT 65535

E30207 Alarm limits permanently violated

E30208 INF_WGT invalid

E30250 Calibration factor missing or incorrect

E30251 Temp. coeff. factors missing or incorrect

E30252 Double fuse defective (zero value)

Scale no. 4

E30253 Temperature sensor missing / defective

Code Description

E30301 Standstill not possible (timeout after 10 sec)

E30302 Weight value out of range (overload or underload)

E30303 Incorrect calibration weight value

E30304 Negative initial weight

E30305 Wrong initial weight

E30306 ACT_WGT 65535

E30307 Alarm limits permanently violated

E30308 INF_WGT invalid

E30350 Calibration factor missing or incorrect

E30351 Temp. coeff. factors missing or incorrect

E30352 Double fuse defective (zero value)

E30353 Temperature sensor missing / defective

2-34 Fresenius Medical Care multiFiltrate TM 6/03.07

Scales nos. 3 and 4 (both)

Chapter 2: Functional Description

Code Description

E30401 Standstill not possible (timeout after 10 sec)

E30402 Weight value out of range (overload or underload)

E30403 Incorrect calibration weight value

E30404 Negative initial weight

E30405 Wrong initial weight

E30406 ACT_WGT 65535

E30407 Alarm limits permanently violated

E30408 INF_WGT invalid

E30450 Calibration factor missing or incorrect

E30451 Temp. coeff. factors missing or incorrect

E30452 Double fuse defective (zero value)

E30453 Temperature sensor missing / defective

2.5.5 Error Codes of Pressures Lo-Level Routines

PART

Code Description

E30551 Calibration factor missing, incorrect

E30552 Offset factor missing, incorrect

E30553 Offset > measurement value

E30557 DAC setting missing, incorrect

PPHF

Code Description

E30651 Calibration factor missing, incorrect

E30652 Offset factor missing, incorrect

E30653 Offset > measurement value

E30657 DAC setting missing, incorrect

Fresenius Medical Care multiFiltrate TM 6/03.07 2-35

Document’s Content and Additional Information Share Manual

Fresenius Medical Care multiFiltrate User manual manual

  • Fresenius Medical Care
  • Medical Equipment
  • User manual manual for Fresenius Medical Care multiFiltrate
  • fresenius-medical-care-multifiltrate-user-manual-manual-282_manual.pdf
  • 282 |

Pages Preview:





Document Transcription:

  • Fresenius Medical Care multiFiltrate Chapter 4: Operation 4-20 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 CRRT treatments 4.5.3.1 Inserting the dialysate or sub predilution tubing system (except for CVVH) 4.5.3.2 Inserting the substituate or sub postdilution tubing system (except for CVVHD)  Insert the dialysate or sub predilution tubing system according to the instructions. …

  • Fresenius Medical Care multiFiltrate Chapter 4: Operation 4-26 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 CRRT treatments 4.5.7 Treatment 4.5.7.1 Treatment main screen The optical detector senses opaque fluid. The blood pump is stopped. If not done yet when using an NaCl solution with two connectors:  Disconnect the venous patient line from the NaCl solution and connect it t …

  • Fresenius Medical Care multiFiltrate Chapter 4: Operation 4-8 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 The target value is set, for example, to 200 ml/min by turning the rotary selector (the target value is shown in white). The selected value must be confirmed by pressing the rotary selector [OK]. After the set value has been confirmed with [OK], the currently selected actua …

See Details
Download

Понравилась статья? Поделить с друзьями:

Это тоже интересно:

  • Freshpack pro вакуумный упаковщик инструкция на русском языке
  • Freemotion b552 инструкция на русском
  • Free buds lite huawei инструкция
  • Freedom soft tampons инструкция по применению на русском
  • Freedconn инструкция на русском t com

  • Подписаться
    Уведомить о
    guest

    0 комментариев
    Старые
    Новые Популярные
    Межтекстовые Отзывы
    Посмотреть все комментарии